TCP303 and TCP404XL Performance Verification
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
5- 33
DC Gain Accuracy
This procedure tests the DC gain accuracy of the TCP303 and TCP404XL. In
this test you compare the voltage output of the amplifier to a reference input.
This test uses a 50-turn current loop that you make using wire and a 3-inch form.
Refer to page 5--4 for instructions on making the current loop.
1.
Using a BNC cable, connect the amplifier
OUTPUT
to the 50
Ω
feed-
through termination. Attach the termination to a BNC-to-dual banana
adapter. Insert the dual-banana adapter into the digital multimeter DC
voltage input, taking care to maintain polarity.
2.
Connect the current loop to the current source as shown in Figure 5--10.
3.
Connect the current probe to the
PROBE INPUT
connector. Do not connect
the probe to the current loop at this time.
Current
source
DMM
+ --
+
--
50 Turns
Current probe
BNC-to-dual
banana adapter
50
Ω
Precision
termination
Output
50
Ω
Coaxial
cable
Current flow
Figure 5- 10: DC gain accuracy test setup for the TCP303 and TCP404XL
Make or check the equipment settings in Table 5--18.
Table 5- 18: Equipment settings for DC gain accuracy
Digital multimeter
Measurement
DC volts
Range
Autoranging
Current source
Output
Off
TCPA300 and TCPA400
Coupling
DC
Equipment Connections
Equipment Settings
Summary of Contents for TCPA300 SERIES
Page 4: ......
Page 20: ...Preface xvi TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Page 21: ...Getting Started...
Page 22: ......
Page 37: ...Operating Basics...
Page 38: ......
Page 45: ...Reference...
Page 46: ......
Page 71: ...Specifications...
Page 72: ......
Page 88: ......
Page 89: ...Performance Verification...
Page 90: ......
Page 135: ...Adjustment Procedures...
Page 136: ......
Page 157: ...Maintenance...
Page 158: ......
Page 181: ...Replaceable Parts...
Page 182: ......
Page 199: ...Glossary and Index...
Page 200: ......