TCP305 and TCP312 Performance Verification
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
5- 23
Table 5- 13: DC gain accuracy test for the TCP305 and TCP312
Probe
Range
Current source
output (ADC)
Expected
output,
V
E
(VAC)
M1 (VDC)
M2 (VDC)
M1- M2 (VDC)
% Error,
calculated
TCP305 5 A/V
±
1.00 A
±
1.00 V
/
±
3.00 A
±
3.00 V
±
5.00 A
±
5.00 V
10 A/V
±
2.00 A
±
1.00 V
/
±
6.00 A
±
3.00 V
±
10.00 A
±
5.00 V
TCP312 1 A/V
±
0.20 A
±
1.00 V
/
±
0.60 A
±
3.00 V
±
1.00 A
±
5.00 V
10 A/V
±
1.20 A
±
0.60 V
/
±
3.60 A
±
1.80 V
±
6.00 A
±
3.00 V
g.
Verify that the measured output (M1--M2) is within the warranted
specification listed in Table 4--1 on page 4--1, by computing %Error as
follows:
%
Error
=
M
1
−
M
2
−
2(
V
E
)
2(
V
E
)
×
100
For example, you may measure values of 0.990V for M1 and --0.998V
for M2. With an expected V
E
of 1.0V, you compute the %Error as
follows:
%
Error
=
0.990
V
−
--0.998
V
−
2(1
V
)
2(1
V
)
×
100
=
1.988
V
−
2
V
2
V
×
100
=
--0.6%
5.
Record the results on the test record.
6.
Disconnect the DMM cable from the amplifer and the probe from the 5T
coil.
Summary of Contents for TCPA300 SERIES
Page 4: ......
Page 20: ...Preface xvi TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Page 21: ...Getting Started...
Page 22: ......
Page 37: ...Operating Basics...
Page 38: ......
Page 45: ...Reference...
Page 46: ......
Page 71: ...Specifications...
Page 72: ......
Page 88: ......
Page 89: ...Performance Verification...
Page 90: ......
Page 135: ...Adjustment Procedures...
Page 136: ......
Page 157: ...Maintenance...
Page 158: ......
Page 181: ...Replaceable Parts...
Page 182: ......
Page 199: ...Glossary and Index...
Page 200: ......