TCP305 and TCP312 Performance Verification
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
5- 29
TCP305 Current Probe Test Record
Photocopy this form and use it to record the performance test results.
TCP305 test record
Instrument Serial Number:
Certificate Number:
Temperature:
Relative Humidity %:
Date of Calibration:
Technician:
Performance test
Range, A/V
Test
current, A
Minimum
Measured/calculated
Maximum
DC Gain Accuracy (% Error)
5 A/V
1.00 A
--3%
+3%
/
3.00 A
--3%
+3%
5.00 A
--3%
+3%
10 A/V
2.00 A
--3%
+3%
/
6.00 A
--3%
+3%
10.00 A
--3%
+3%
Rise Time
5 A/V
------
≤
7 ns
Bandwidth
5 A/V
.707
------------