Reference Notes
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
3- 5
0A
Maximum
pulsed
current
Maximum
continuous
current
50%
Pulse width
at 50%
50%
Do Not Exceed
Pulse width
at 50%
I
max
p
I
max
c
(a) Maximum allowable pulse width
(b) Maximum allowable pulse amplitude
Figure 3- 3: Applying the amp-second product rule
To determine the maximum allowable pulse amplitude, measure the pulse width
at the 50% points (see Figure 3--3b). Divide the Ampere-second (or Ampere-
microsecond) specification of your probe by the pulse width. The quotient is the
maximum allowable current; the peak amplitude of the measured pulse must be
less than this value.
For example, the TCP312 Current Probe has a maximum Ampere-second
product of 500 A
⋅
s in the 10 A/V range setting. If a pulse measured with a
TCP312 probe has a width of 15
s, the maximum allowable peak current would
be 500 A
⋅
s divided by 15
s, or 33.3 A.
Procedure B
Summary of Contents for TCPA300 SERIES
Page 4: ......
Page 20: ...Preface xvi TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Page 21: ...Getting Started...
Page 22: ......
Page 37: ...Operating Basics...
Page 38: ......
Page 45: ...Reference...
Page 46: ......
Page 71: ...Specifications...
Page 72: ......
Page 88: ......
Page 89: ...Performance Verification...
Page 90: ......
Page 135: ...Adjustment Procedures...
Page 136: ......
Page 157: ...Maintenance...
Page 158: ......
Page 181: ...Replaceable Parts...
Page 182: ......
Page 199: ...Glossary and Index...
Page 200: ......