TCPA300 and TCPA400 Performance Verification
5- 16
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
TCPA300 Amplifier Test Record
Photocopy this form and use it to record the performance test results.
TCPA300 amplifier test record
Instrument Serial Number:
Certificate Number:
Temperature:
Relative Humidity %:
Date of Calibration:
Technician:
Performance test
Range, A/V
Minimum
Measured/calculated
Maximum
Front-Panel Display
Pass/Fail
DC Gain Accuracy (% Error)
1
--1%
+1%
5
--1%
+1%
10
--1%
+1%
50
--1%
+1%
50 (COMP)
--1%
+1%
Bandwidth
1
.707
------------
1 (100 MHz)
.707
------------
5
.707
------------
10
.707
------------
50
.707
------------
50 (COMP)
.707
------------
AC Coupling
1
TEKPROBE Cable
1
1.05 Ap-p
1.5 Ap-p
BNC Coax Cable
1
1.05 Vp-p
1.5 Vp-p
Degauss
Pass/Fail
Current Overload
Pass/Fail
1
The AC Coupling test can be performed with either cable, but the units of measure for the results differ between the two.
Summary of Contents for TCPA300 SERIES
Page 4: ......
Page 20: ...Preface xvi TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Page 21: ...Getting Started...
Page 22: ......
Page 37: ...Operating Basics...
Page 38: ......
Page 45: ...Reference...
Page 46: ......
Page 71: ...Specifications...
Page 72: ......
Page 88: ......
Page 89: ...Performance Verification...
Page 90: ......
Page 135: ...Adjustment Procedures...
Page 136: ......
Page 157: ...Maintenance...
Page 158: ......
Page 181: ...Replaceable Parts...
Page 182: ......
Page 199: ...Glossary and Index...
Page 200: ......