Specifications
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
4- 11
TCP303 Max peak pulse vs PW
0
100
200
300
400
500
600
Am
peres
(P
eak)
0
50
100
150
50 A/V
5 A/V
Allowable pulse width (
µ
seconds)
0
10
20
30
40
50
60
Am
peres
(P
eak)
0
5
10
15
20
0
10
20
30
40
50
60
Am
peres
(P
eak)
0
TCP312 Max peak pulse vs PW
TCP305 Max peak pulse vs PW
5
10
15
20
25
≤
25 A continuous
≤
30 A continuous
15000 A*
µ
s
3000 A*
µ
s
Any width
Any width
10 A/V
1 A/V
Any width
≤
25A continuous
Allowable pulse width (
µ
seconds)
Allowable pulse width (
µ
seconds)
Maximum peak pulse
≤
500A
≤
5 A continuous
500 A*
µ
s
500 A*
μ
s
50 A*
µ
s
Maximum peak pulse
≤
50A
25
Maximum peak pulse
≤
50A
Range setting
≤
212 A peak (
≤
150 A RMS continuous)
10 A/V
5 A/V
Range setting
Range setting
Figure 4- 7: Specified operating area of the probes
Summary of Contents for TCPA300 SERIES
Page 4: ......
Page 20: ...Preface xvi TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Page 21: ...Getting Started...
Page 22: ......
Page 37: ...Operating Basics...
Page 38: ......
Page 45: ...Reference...
Page 46: ......
Page 71: ...Specifications...
Page 72: ......
Page 88: ......
Page 89: ...Performance Verification...
Page 90: ......
Page 135: ...Adjustment Procedures...
Page 136: ......
Page 157: ...Maintenance...
Page 158: ......
Page 181: ...Replaceable Parts...
Page 182: ......
Page 199: ...Glossary and Index...
Page 200: ......