Disassembly
7- 18
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
TCP303 and TCP404XL
The following procedures explain how to disassemble the probe body and
replace the current transformer, the circuit board, and the cable assembly.
WARNING.
Probe disassembly must be performed by qualified service personnel.
1.
Unlock the probe slide.
2.
Place the probe on a flat surface with the screw heads facing up.
3.
Using a
3
/
32
inch Allen wrench, remove the eight retaining screws from the
case.
4.
While holding the squeeze handle and slide, lift the top part of the probe
body off.
5.
Remove the slide.
WARNING.
There is a spring in the squeeze handle that can pop out and cause
personal injury unless care is exercised when disassembling the probe.
6.
Lift the cable out of the probe handle and insert a pair of pliers in the handle
as shown in Figure 7--15.
Body half
Squeeze
handle
Torsion spring
Figure 7- 15: Removing the handle
7.
While maintaining a firm grip on the pliers, carefully lift the squeeze handle
and gear out of the probe. See Figures 7--15 and 7--16.
Summary of Contents for TCPA300 SERIES
Page 4: ......
Page 20: ...Preface xvi TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Page 21: ...Getting Started...
Page 22: ......
Page 37: ...Operating Basics...
Page 38: ......
Page 45: ...Reference...
Page 46: ......
Page 71: ...Specifications...
Page 72: ......
Page 88: ......
Page 89: ...Performance Verification...
Page 90: ......
Page 135: ...Adjustment Procedures...
Page 136: ......
Page 157: ...Maintenance...
Page 158: ......
Page 181: ...Replaceable Parts...
Page 182: ......
Page 199: ...Glossary and Index...
Page 200: ......