Disassembly
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
7- 13
Probes
Use these procedures when repairing or replacing defective components in the
current probes.
WARNING.
Probe disassembly must be performed by qualified service personnel.
The probes that mate with the TCPA300 and TCPA400 Current Probe Amplifiers
have two physical styles. The TCP305 and TCP312 use a slender form factor
that contain three replaceable assemblies -- the connector/cable, the circuit board,
and the current transformer.
The TCP303 and TCP404XL use a pistol-style case with an integral handgrip.
These probes contain two replaceable assemblies -- the connector/cable/circuit
board assembly and the current transformer. Both probes use a few mechanical
components that are replaceable, such as springs and plastic parts.
Equipment Required
Table 7--2 lists the equipment required to service the probes.
Table 7- 2: Equipment required
Qty
Description
1
#2 Phillips screwdriver
1
Isopropyl alcohol
1
Silicone-based grease
1
25 W soldering iron
Summary of Contents for TCPA300 SERIES
Page 4: ......
Page 20: ...Preface xvi TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Page 21: ...Getting Started...
Page 22: ......
Page 37: ...Operating Basics...
Page 38: ......
Page 45: ...Reference...
Page 46: ......
Page 71: ...Specifications...
Page 72: ......
Page 88: ......
Page 89: ...Performance Verification...
Page 90: ......
Page 135: ...Adjustment Procedures...
Page 136: ......
Page 157: ...Maintenance...
Page 158: ......
Page 181: ...Replaceable Parts...
Page 182: ......
Page 199: ...Glossary and Index...
Page 200: ......