TCPA300 and TCPA400 Performance Verification
5- 8
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
d.
Calculate the %Error between the measured output, M
1
, and the
expected output (V
E
) %Error as follows:
%
Error
=
M
1
−
V
E
V
E
×
100
For example, using an expected output
V
E
of 2.5000 V and a measured
output
M
1
of 2.510, the
%Error
would be:
%
Error
=
2.5100
−
2.5000
2.5000
×
100
= +
0.4%
NOTE
. It is important that the correct polarity be used to calculate the %Error.
6.
Verify that the measured DC gain accuracy is within the limits specified for
all settings in the test record on page 5--16 for the TCPA300 (page 5--17 for
the TCPA400). Record the %Error in the test record.
NOTE
. If the %Error on any of the settings is greater than the specified limit, you
must perform the adjustment procedure that begins on page 6--3.
7.
Disconnect the DMM cable from the amplifier and the calibration adapter
from the current source.
Summary of Contents for TCPA300 SERIES
Page 4: ......
Page 20: ...Preface xvi TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Page 21: ...Getting Started...
Page 22: ......
Page 37: ...Operating Basics...
Page 38: ......
Page 45: ...Reference...
Page 46: ......
Page 71: ...Specifications...
Page 72: ......
Page 88: ......
Page 89: ...Performance Verification...
Page 90: ......
Page 135: ...Adjustment Procedures...
Page 136: ......
Page 157: ...Maintenance...
Page 158: ......
Page 181: ...Replaceable Parts...
Page 182: ......
Page 199: ...Glossary and Index...
Page 200: ......