TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
7- 21
Isolating Hardware Faults
Use the following procedures to help locate a hardware malfunction.
You can isolate the symptoms to the amplifier or probe by substituting a known
good probe or amplifier. If you suspect that you have a defective probe and you
don’t have another compatible probe, you can use the calibration adapter in place
of the probe to verify some of the amplifier functions.
Tables 7--3 and 7--4 list some of the failures that can occur with the amplifiers
and probes.
Table 7- 3:
Amplifier
hardware faults
Symptom
Possible cause
Repair action
Amplifier does not power on. LEDs do not
turn on and fan does not work.
Internal power supply problem
Connector loose between main board and
front panel
Front panel is defective
Check for +15 V on power supply
connector J2, pin 1 at main board. If OK,
replace amplifier board. If not, replace
power supply.
Secure ribbon cable from front-panel
assembly to main board
Replace front-panel assembly
LEDs do not turn on and fan works.
Internal power supply problem
Connector loose between main board and
front panel
Front panel is defective
Check for +15 V on power supply
connector J2, pin 1 at main board. If OK,
check power supply test points on main
board: +5VD, +5VA, --5VA, +10V, --10V.
If these voltages check OK, replace the
front-panel assembly. If not, replace the
main board.
Secure ribbon cable from front-panel
assembly to main board
Replace front-panel assembly
LEDs turn on and fan does not work.
Defective fan or defective power supply
Check for greater than +7V at fan
connector J3, pin 1 at main board. If
present, replace fan. If not, check for +15
V input from power supply. If +15 V is
present, replace the amplifier board. If
not, replace the power supply.
LEDs turn on in a random pattern
(different than is documented in this
manual)
Front panel is defective
Replace front-panel assembly
One amplifier range does not work, other
range is OK. (TCPA300 only)
Defective amplifier circuit
Replace main board assembly
Summary of Contents for TCPA300 SERIES
Page 4: ......
Page 20: ...Preface xvi TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Page 21: ...Getting Started...
Page 22: ......
Page 37: ...Operating Basics...
Page 38: ......
Page 45: ...Reference...
Page 46: ......
Page 71: ...Specifications...
Page 72: ......
Page 88: ......
Page 89: ...Performance Verification...
Page 90: ......
Page 135: ...Adjustment Procedures...
Page 136: ......
Page 157: ...Maintenance...
Page 158: ......
Page 181: ...Replaceable Parts...
Page 182: ......
Page 199: ...Glossary and Index...
Page 200: ......