2260B Series User Manual
62
OVP Level
3.
Use the Voltage knob to set the
OVP level.
Voltage
OCP Level
4.
Use the Current knob to set the
OCP level.
Current
5.
Press OVP/OCP again to exit. The
OVP/OCP indicator will turn off.
Power switch trip
Set F-95 (Power switch trip) to 1 (to
disable the power switch trip) or to
0 (to enable the power switch trip)
and save.
Page 120
F-95
1 (Disable) or 0 (Enable)
Clear OVP/OCP
protection
The OVP or OCP protection can be
cleared after it has been tripped by
holding the OVP/OCP button for 2
seconds.
(Only applicable when the power
switch trip setting is disabled
[F-95 = 1])
(hold)