6.
Use the current knob to set the
Password for the F-00 setting. The
Password is
“
5 0 0 4
”
.
Current
7.
Press the Voltage knob to enter the
Special Mode. SF- 1 will be
displayed when successful.
Voltage
X can be anything in the display
8.
Rotate the voltage knob to change
the SF setting to SF- 2.
Voltage
9.
The display will show SF- 2
on the top.
10.
Use the current knob to set SF- 2 to
“
1
”
.
Current
11.
Press the Voltage knob to copy the
new firmware to the 2260B.
Voltage