APPENDIX
193
General Specifications
Weight (main unit only)
kg
Approx. 7.5kg
Dimensions (WxHxD)
mm
3
214×124×350
Cooling
Forced air cooling by internal fan.
EMC
Complies with the European EMC directive
2004/108/EC for Class A test and
measurement products.
Safety
Complies with the European Low Voltage
Directive 2006/95/EC and carries the CE-
marking.
Withstand voltage
Between input and chassis: No abnormalities
at 1500 Vac for 1 minute.
Between input and output: No abnormalities
at 3000 Vac for 1 minute.
Between output and chassis: No
abnormalities at 500 Vdc for 1 minute for
30V, 80V models.
No abnormalities at 1500 Vdc for 1 minute
for 250V, 800V models.
Insulation resistance
Between input and chassis: 500 Vdc, 100M
Ω
or more
Between input and output: 500 Vdc, 100M
Ω
or more
Between output and chassis: 500 Vdc,
100M
Ω
or more for 30V, 80V and 250V
models.
1000Vdc, 100M
Ω
or more for 800V models.
*1: At 85 ~ 132Vac or 170 ~ 265Vac, constant load.
*2: From No-load to Full-load, constant input voltage. Measured at the sensing
point in Remote Sense.
*3: Measure with JEITA RC-9131B (1:1) probe
*4: Measurement frequency bandwidth is 10Hz to 20MHz.
*5: Measurement frequency bandwidth is 5Hz to 1MHz.
*6: From 10% to 90% of rated output voltage, with rated resistive load.
*7: From 90% to 10% of rated output voltage, with rated resistive load.
*8: Time for output voltage to recover within 0.1% + 10mV of its rated output
for a load change from 50 to 100% of its rated output current.
*9: For load voltage change, equal to the unit voltage rating, constant input
voltage.
Summary of Contents for Keithley 2260B Series
Page 14: ...2260B Series User Manual 10 Considerations 33 Grounding 36 ...
Page 43: ...OPERATION 39 Remove Test Sequence 99 Test Sequence Error Messages 100 ...
Page 52: ...2260B Series User Manual 48 Removal Reverse the procedure to remove the terminal covers ...
Page 111: ...CONFIGURATION 107 Note Power On and Calibration settings can only be set during power up ...