ANALOG CONTROL
135
Connection
2260B
Switch
Analog
connector
24
2
Output
Terminal
2 core shielded
wire or twisted
pair
Pin2 → Switch
Pin24
→ Switch
Wire shield → negative (
-) output terminal
Steps
1.
Connect the external switch according to the
connection diagrams above.
Set F-94 (External output logic) in
the power on configuration
settings to 0 (High = On) or 1
(Low = On).
Be sure to cycle the power after
setting the power on configuration
settings.
Page 120
2.
Press the Function key and confirm
the new configuration settings.
3.
The switch is now ready to set the output on or
off.