Setting Power On Configuration Settings
Background
The Power On configuration settings can only
be changed during power up to prevent the
configuration settings being inadvertently
changed.
Ensure the load is not connected.
Ensure the power supply is off.
Steps
1.
Hold the Function key while
turning the power on.
2.
The display will show F-90 on the
top and the configuration setting
for F-90 on the bottom.
3.
Rotate the Voltage knob to change
the F setting.
Voltage
Range
F-90~ F-95
4.
Use the Current knob to set the
parameter for the chosen F setting.
Current