OPERATION
61
ALM
V
A
Before setting the OVP or OCP level:
Ensure the load is not connected.
Ensure the output is set to off.
Setting Ranges
2260B (360W)
30-36
80-13
250-4
800-1
OVP Range (V)
3-33
8-88
20-275
20-880
OCP Range (A)
3.6-39.6
1.35-14.85 0.45-4.95
0.144-1.584
2260B (720W)
30-72
80-27
250-9
800-2
OVP Range (V)
3-33
8-88
20-275
20-880
OCP Range (A)
5-79.2
2.7-29.7
0.9-9.9
0.288-3.168
2260B (1080W)
30-108
80-40
250-13
800-4
OVP Range (V)
3-33
8-88
20-275
20-880
OCP Range (A)
5-118.8
4.05-44.55 1.35-14.85 0.432-4.752
Steps
1.
Press the OVP/OCP key. The
OVP/OCP key lights up.
2.
The OVP setting will be displayed on the top
and the OCP setting (or OFF) will be displayed
on the bottom.
V
A
OVP Setting
OCP Setting