Cycle Step Start: Determines the starting step for the Cycle
Repeat field. Range: 1~999. The Cycle Step Start field must be less
than the Cycle Step End field.
Cycle Step End: Determines the ending step for the Cycle Repeat
field. Range: 2~1000. The Cycle Step End field must be greater
than the Cycle Step Start field.
Line 2:
Step Header: This field contains the “Step” header text. Required.
Do not remove from this location.
Time(sec) Header: Th
is field contains the “Time(sec)” header text.
Required. Do not remove or alter from this location.
Voltage (V) Header: This field contains the “Voltage (V)” header
text. Required. Do not remove or alter from this location.
Current (A) Header: This field c
ontains the “Current (A)” header
text. Required. Do not remove or alter from this location.
Line 3+:
Step Number: The Step Number field numbers each step in the
test function. It has a range of 1~1000. All steps must be
numbered in numerical order. No step
number can be “skipped”
between the first and last step.
Time(sec): Designates the duration time of the corresponding
step in seconds. It has a range of 0.05 seconds to 20 days. It has a
step resolution of 0.05 seconds (50ms). A setting of 0 will skip the
step.
Voltage (V): Designates the voltage of the corresponding steps in
volts. MIN or MAX can be used to set the voltage to minimum or
maximum, respectively.
Current (A): Designates the current of the corresponding steps in
amps. MIN or MAX can be used to set the current to minimum or
maximum, respectively.
Summary of Contents for Keithley 2260B Series
Page 14: ...2260B Series User Manual 10 Considerations 33 Grounding 36 ...
Page 43: ...OPERATION 39 Remove Test Sequence 99 Test Sequence Error Messages 100 ...
Page 52: ...2260B Series User Manual 48 Removal Reverse the procedure to remove the terminal covers ...
Page 111: ...CONFIGURATION 107 Note Power On and Calibration settings can only be set during power up ...