Example:
File: T001.csv
In the example below there are 10 steps in total.
As designated by the Cycle fields, steps 3~5 are
repeated twice.
In total the steps will run as follows:
Steps 3~5 are repeated twice
Step: 1>2>3>4>5>3>4>5>6>7>8>9>10
Cycle,2,3,5
Step,Time(sec),Voltage (V),Current (A)
1,1,MIN,MIN
2,1,0.05,0.05
3,1,0.1,0.1
4,1,0.15,0.15
5,1,0.2,0.2
6,1,0.25,0.25
7,1,0.3,0.3
8,1,0.35,0.35
9,1,0.4,0.4
10,1,0.45,0.45