OPERATION
73
Serial 2260B
Units
1.
a. Connect the 1
st
Sense+ terminal to the
positive potential of the load.
b. Connect the 1
st
Sense- terminal to the
positive output terminal of the second 2260B
unit.
c. Connect the 2
nd
Sense+ terminal to the
positive terminal of the second 2260B unit.
d. Connect the 2
nd
Sense- terminal to negative
terminal of the load.
2260B #2
a
b
c
d
Load
Input
Input
Output
Sense
Sense
2260B #1
Output
Output
Sense
Sense
Output
Page 45
2.
Operate the instrument as normal.
See the Serial Operation chapter
for details.