2260B Series User Manual
82
Note
Configuration settings can be checked for both the
master and slave units by pressing the Function
key and checking F-93.
Only the Master OVP and OCP level is used for
over voltage and current protection. Slave OVP and
OCP level is disregarded.
OTP works independently for each unit.
Master-Slave
Operation
Only operate the power supplies in parallel if
the units are configured correctly.
1.
Turn on the master and slave units. The slave
unit(s) will show a blank display.
Master unit
Slave units
2.
Operation of all units is controlled
via the master unit. Operation of
the master unit is the same as for a
single unit. See the Basic Operation
chapter.
Page 60.
3.
Press the Output key to begin.
Caution
Only operate the power supplies in parallel if using
units of the same model number.
Only a maximum of 3 units can be used in parallel.