APPENDIX
181
200Vac
A
2.5
Inrush current
Less than 25A.
Maximum input power
VA
500
Power factor
100Vac
0.99
200Vac
0.97
Efficiency
100Vac
%
77
78
79
80
200Vac
%
79
80
81
82
Hold-up time
20ms or greater
Interface Capabilities
USB
TypeA: Host, TypeB: Slave, Speed: 1.1/2.0,
USB Class: CDC(Communications Device
Class)
LAN
MAC Address, DNS IP Address, User
Password, Gateway IP Address, Instrument
IP Address, Subnet Mask
GPIB
Optional: GUG-001 (GPIB to USB Adapter)
Environmental Conditions
Operating temperature
0
˚
C to 50
˚
C
Storage temperature
-25
˚
C to 70
˚
C
Operating humidity
20% to 85% RH; No condensation
Storage humidity
90% RH or less; No condensation
Altitude
Maximum 2000m
General Specifications
Weight (main unit only)
kg
Approx. 3kg
Dimensions (WxHxD)
mm
3
71×124×350
Cooling
Forced air cooling by internal fan.
EMC
Complies with the European EMC directive
2004/108/EC for Class A test and
measurement products.
Safety
Complies with the European Low Voltage
Directive 2006/95/EC and carries the CE-
marking.
Withstand voltage
Between input and chassis: No
abnormalities at 1500 Vac for 1 minute.
Between input and output: No
abnormalities at 3000 Vac for 1 minute.
Between output and chassis: No
abnormalities at 500 Vdc for 1 minute for
30V and 80V models.
No abnormalities at 1500 Vdc for 1 minute
for 250V, 800V models.
Summary of Contents for Keithley 2260B Series
Page 14: ...2260B Series User Manual 10 Considerations 33 Grounding 36 ...
Page 43: ...OPERATION 39 Remove Test Sequence 99 Test Sequence Error Messages 100 ...
Page 52: ...2260B Series User Manual 48 Removal Reverse the procedure to remove the terminal covers ...
Page 111: ...CONFIGURATION 107 Note Power On and Calibration settings can only be set during power up ...