CONFIGURATION
109
V-I Mode
Selects High Speed Priority or Slew Rate
Priority for CV or CC mode. The voltage or
current slew rate can only be edited if CC/CV
Slew Rate Priority is selected. The ISR indicator
will be lit for CC Slew Rate Priority and the
VSR indicator will be lit for CV Slew Rate
Priority.
Note: CC and CV Slew Rate Priority mode are
disabled when voltage/current output is set to
external control.
ISR
CC Slew Rate priority
VSR
CV Slew Rate priority
F-03
0 = CV high speed priority
1 = CC high speed priority
2 = CV slew rate priority
3 = CC slew rate priority
Rising Voltage
Slew Rate
Sets the rising voltage slew rate. Only
applicable if V-I Mode is set to CV Slew Rate
Priority.
F-04
0.01V/s~60V/s (2260B-30-XX)
0.1V/s~160V/s (2260B-80-XX)
0.1V/s~500.0V/s (2260B-250-XX)
1V/s~1600V/s (2260B-800-XX)
Falling Voltage
Slew Rate
Sets the falling voltage slew rate. Only
applicable if V-I Mode is set to CV Slew Rate
Priority.
F-05
0.01V/s~60V/s (2260B-30-XX)
0.1V/s~160V/s (2260B-80-XX)
0.1V/s~500.0V/s (2260B-250-XX)
1V/s~1600V/s (2260B-800-XX)
Summary of Contents for Keithley 2260B Series
Page 14: ...2260B Series User Manual 10 Considerations 33 Grounding 36 ...
Page 43: ...OPERATION 39 Remove Test Sequence 99 Test Sequence Error Messages 100 ...
Page 52: ...2260B Series User Manual 48 Removal Reverse the procedure to remove the terminal covers ...