Commands Listed in Alphabetical Order
pane of the Measurement Setup dialog box. (If tracking is enabled for Low, the
instrument automatically calculates the low value using the method set with the
command, MEASUrement:MEAS<x>:SOURCE<x>:HILow:METHod.)
NOTE.
The SOURCE<x> argument speci
fi
es the source, which can be source 1
or source 2. To set the actual waveform (channel, reference, or math) that is
assigned to source, you must use the MEASUrement:MEAS<x>:SOURCE<x>
WFM command. Two sources are provided because some measurements (gain,
phase) require two sources.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:SOUrce<x>:HILow:TRACk:LOW:VALue <NR3>
MEASUrement:MEAS<x>:SOUrce<x>:HILow:TRACk:LOW:VALue?
Related Commands
MEASUrement:MEAS<x>:SOUrce<x>:HILow:METHod
MEASUrement:MEAS<x>:SOUrce<x>:HILow:TRACk:LOW:ENABle
Arguments
NR3
is the absolute value of the low level for the HiLow settings.
Examples
MEASUREMENT:MEAS2:SOURCE1:HILOW:TRACK:LOW:VALUE 5.0E-3
sets the
low value used to take Measurement 2 on the Source 1 waveform to 5 mV.
MEASUREMENT:MEAS3:SOURCE1:HILOW:TRACK:LOW:VALUE?
might
return
MEASUREMENT:MEAS3:SOURCE1 HILOW:TRACK:LOW:VALUE
1.00000000000E-002
, indicating that the low value used to take Measurement
3 on the Source 1 waveform is 10 mV.
MEASUrement:MEAS<x>:SOUrce<x>:WFM
This command sets or queries the source waveform (SOURCE1) on which the
speci
fi
ed single-channel measurements are taken and
from which
two-channel
measurements, such as delay or phase are taken. The measurement slot is speci
fi
ed
by x, and can be 1 through 8. This command also sets or queries the destination
waveform (SOURCE2)
to which
two-channel measurements are taken. If the
source is not speci
fi
ed (either 1 or 2), the default is SOURCE1. Optionally,
this command can specify the timebase for the speci
fi
ed source or destination
waveform. If the timebase is not speci
fi
ed, the default is the Main timebase.
Group
Measurement
DSA8300 Programmer Manual
2-257
Summary of Contents for DSA8300
Page 1: ...xx DSA8300 Digital Serial Analyzer ZZZ Programmer Manual P077057002 077 0570 02 ...
Page 2: ......
Page 8: ...Preface iv DSA8300 Programmer Manual ...
Page 9: ...Getting Started ...
Page 10: ......
Page 15: ...Syntax and Commands ...
Page 16: ......
Page 404: ...Commands Listed in Alphabetical Order 2 388 DSA8300 Programmer Manual ...
Page 405: ...Status and Events ...
Page 406: ......
Page 426: ...Status and Events 3 20 DSA8300 Programmer Manual ...
Page 427: ...Examples ...
Page 428: ......
Page 431: ...Appendices ...
Page 432: ......
Page 433: ...Appendix A Character Set DSA8300 Programmer Manual A 1 ...
Page 434: ...Appendix A Character Set A 2 DSA8300 Programmer Manual ...
Page 446: ...Appendix C Factory Default Setup Values C 8 DSA8300 Programmer Manual ...