Commands Listed in Alphabetical Order
Arguments
EDGE
sets subsequent autoset commands to con
fi
gure the instrument to
display a rising edge of the signal connected to the currently selected channel
in the center 20% of the graticule.
PERIod
sets subsequent autoset commands to con
fi
gure the instrument to
display 2 to 3 periods of the signal connected to the currently selected channel
centered in the graticule.
EYE
sets subsequent autoset commands to con
fi
gure the instrument to display
one bit (two eye crossings) horizontally centered across approximately 70%
of the graticule (7 divisions) and vertically centered across approximately
60% of the graticule (6 divisions). Use the Eye argument for N
RZ
waveforms.
RZ
sets subsequent autoset commands to con
fi
gure the instrument to display
one bit (two rising edges) horizontally centered across approximately 60%
of the graticule (6 divisions) and vertically centered across approximately
50% of the graticule (5 divisions).
TDR
sets subsequent autoset commands to con
fi
gure the instrument to display
the
fi
rst signi
fi
cant re
fl
ection at the third horizontal division. If the Display
Incident Edge option is on, the incident edge is placed at the one-half
horizontal division (from left edge). If the Display Incident Edge is off,
the incident edge is placed one horizontal division off the left edge of the
screen. If no signi
fi
cant re
fl
ections are located, the incident edge is placed
at the one-half horizontal division (from left edge) regardless of the Display
Incident Edge setting.
NOTE.
If the
Trigger Source
is a clock and the trigger Mode is set to Pattern, an
Edge or Period autoset will attempt to display 20 bit periods.
Examples
AUTOSET:TYPE EDGE
sets the default Autoset mode to EDGE.
AUTOSET:TYPE?
might return
AUTOSET:TYPE RZ
, indicating that the default
Autoset mode is RZ.
AUTOSet:UNDO (No Query Form)
This command (no query form) returns the parameters changed by an AUTOSet
command to their previous states. This command is invalidated the
fi
rst time you
modify any instrument state parameter.
Group
Miscellaneous
Syntax
AUTOSet:UNDO
DSA8300 Programmer Manual
2-71
Summary of Contents for DSA8300
Page 1: ...xx DSA8300 Digital Serial Analyzer ZZZ Programmer Manual P077057002 077 0570 02 ...
Page 2: ......
Page 8: ...Preface iv DSA8300 Programmer Manual ...
Page 9: ...Getting Started ...
Page 10: ......
Page 15: ...Syntax and Commands ...
Page 16: ......
Page 404: ...Commands Listed in Alphabetical Order 2 388 DSA8300 Programmer Manual ...
Page 405: ...Status and Events ...
Page 406: ......
Page 426: ...Status and Events 3 20 DSA8300 Programmer Manual ...
Page 427: ...Examples ...
Page 428: ......
Page 431: ...Appendices ...
Page 432: ......
Page 433: ...Appendix A Character Set DSA8300 Programmer Manual A 1 ...
Page 434: ...Appendix A Character Set A 2 DSA8300 Programmer Manual ...
Page 446: ...Appendix C Factory Default Setup Values C 8 DSA8300 Programmer Manual ...