Commands Listed in Alphabetical Order
Examples
MEASUREMENT:MEAS2:SOURCE1:EDGE:DIRECTION FORWARD
causes the
instrument to search forward from the beginning of the Source 1 waveform (or
the beginning of the gated region, if gates are enabled) for the
fi
rst speci
fi
ed edge
(rising or falling).
MEASUREMENT:MEAS8:SOURCE1:EDGE:DIRECTION?
might return
MEASUREMENT:MEAS8:SOURCE1:EDGE:DIRECTION FORWARD
, indicating that
the instrument will search forward from the beginning of the Source 1 waveform
(or the beginning of the gated region, if gates are enabled) for the speci
fi
ed edge
(rising or falling) when taking Measurement 8.
MEASUrement:MEAS<x>:SOUrce<x>:EDGE:SLOPe
This command sets or queries the slope (or polarity) of the edges used in delay
time measurements. The measurement slot is speci
fi
ed by x, which can be 1
through 8; the source is speci
fi
ed by x, which can be 1 (source waveform) or
2 (destination waveform). For edge measurements other than delay, the slope
is determined implicitly (for example, rise time measurements use positive
slopes). If gates are enabled (see the
MEASUrement:MEAS<x>:GATing:STATE
command), the instrument
fi
nds the edge within the measurement zone; otherwise,
the instrument
fi
nds the edge within the entire waveform record.
Sending this command is equivalent to selecting an option button (±, +, or -) for
Slope in the Edges section of the Region dialog in the Measurement Setup dialog
box. (A delay measurement must be selected before these buttons are active.)
Use the MEASUrement:MEAS<x>:SOURCE:EDGE:DIRection command to
set the direction that the instrument uses to look for the rising or falling edge
of the waveforms.
NOTE.
The SOURCE<x> argument speci
fi
es the source, which can be source 1
or source 2. To set the actual waveform (channel, reference, or math) that is
assigned to source, you must use the MEASUrement:MEAS<x>:SOURCE<x>
WFM command. Two sources are provided because some measurements (such as
gain and phase) require two sources.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:SOUrce<x>:EDGE:SLOPe {RISE | FALL |
EITHer}
MEASUrement:MEAS<x>:SOUrce<x>:EDGE:SLOPe?
DSA8300 Programmer Manual
2-247
Summary of Contents for DSA8300
Page 1: ...xx DSA8300 Digital Serial Analyzer ZZZ Programmer Manual P077057002 077 0570 02 ...
Page 2: ......
Page 8: ...Preface iv DSA8300 Programmer Manual ...
Page 9: ...Getting Started ...
Page 10: ......
Page 15: ...Syntax and Commands ...
Page 16: ......
Page 404: ...Commands Listed in Alphabetical Order 2 388 DSA8300 Programmer Manual ...
Page 405: ...Status and Events ...
Page 406: ......
Page 426: ...Status and Events 3 20 DSA8300 Programmer Manual ...
Page 427: ...Examples ...
Page 428: ......
Page 431: ...Appendices ...
Page 432: ......
Page 433: ...Appendix A Character Set DSA8300 Programmer Manual A 1 ...
Page 434: ...Appendix A Character Set A 2 DSA8300 Programmer Manual ...
Page 446: ...Appendix C Factory Default Setup Values C 8 DSA8300 Programmer Manual ...