Commands Listed in Alphabetical Order
Examples
MASK?
might return
MASK:AUTOS:MOD MAN; HIL:METH MEAN; :MASK:STAN
OC12; COLO 16711935; COUN:STATE 1; TOT 617435; WAV 3811;SAMP
1905500; MASK:DIS 1;MAR:PERC 0.0000;STATE 1; MASK:MASK1:COUN
0; NR_P 4; POI 20.69040050401E–9,0.00100000005,
20.69499949187E–9,799.99997979030E–6,
22.30039974904E–9,799.99997979030E–6,
22.30500051326E–9,0.00100000005;
POINTSP14.8296,0.0000,15.0300,10.0250,
84.9699,10.0250,85.1703,0.0000; MASK:MASK2:COUN
617435;NR_P 6; POI 21.09520025328E-9,0.0000,
21.33899990042E–9,–340.54054412991E–6,
21.65639934049E–9,–340.54054412991E–6,
21.90020076398E–9,0.0000,
21.65639934049E–9,345.94594035298E–6,
21.33899990042E–9,345.94594035298E–6;
POINTSP 32.4649,50.1253,43.0861,67.1679,
56.9138,67.1679,67.5350,50.1253,56.9138,
32.8320,43.0861,32.8320; MASK:MASK3:COUN
0;NR_P 4; POI 20.69040050401E–9,-0.00100000005,
22.30500051326E–9,–0.00100000005,
22.30039974904E–9,–799.99997979030E–6,
20.69499949187E–9,–799.99997979030E–6;
POINTSP 14.8296,100.2506,85.1703,100.2506,
84.9699,90.2255,15.0300,90.2255; MASK:MASK4:COUN –1;NR_P
0; POI 0,0; POINTSP 0,0;:MASK:MASK5:COUN –1; NR_P 0;POI
0,0;POINTSP 0,0; :MASK:MASK6:COUN –1; NR_P 0;POI 0,0;POINTSP
0,0; :MASK:MASK7:COUN –1; NR_P 0;POI 0,0;POINTSP 0,0;
:MASK:MASK8:COUN –1; NR_P 0;POI 0,0;POINTSP 0,0; :MASK:SOU
CH1,MAIN; WFMDB:STATE 1
.
MASK:AUTOFit EXECute (No Query Form)
This command runs the Auto
fi
t Mask function. The Auto
fi
t Mask function adjusts
the active mask to the source waveform such that the left edge of the mask aligns
with the left-most displayed eye crossing and the mask width is the standard's UI at
the selected horizontal scale. When mask auto
fi
t runs successfully, the instrument
moves the horizontal reference point to line up with the left edge of the mask.
For relative masks, the mask is adjusted vertically so that the mask high and low
positions are aligned with the measured high and low of the waveform.
For absolute masks, the mask is adjusted vertically so that the mask high and
low positions are based on the vertical scale, offset and position of the mask
source waveform.
Running a mask auto
fi
t turns off mask margin (if it is on) but does not change
the percentage value.
DSA8300 Programmer Manual
2-205
Summary of Contents for DSA8300
Page 1: ...xx DSA8300 Digital Serial Analyzer ZZZ Programmer Manual P077057002 077 0570 02 ...
Page 2: ......
Page 8: ...Preface iv DSA8300 Programmer Manual ...
Page 9: ...Getting Started ...
Page 10: ......
Page 15: ...Syntax and Commands ...
Page 16: ......
Page 404: ...Commands Listed in Alphabetical Order 2 388 DSA8300 Programmer Manual ...
Page 405: ...Status and Events ...
Page 406: ......
Page 426: ...Status and Events 3 20 DSA8300 Programmer Manual ...
Page 427: ...Examples ...
Page 428: ......
Page 431: ...Appendices ...
Page 432: ......
Page 433: ...Appendix A Character Set DSA8300 Programmer Manual A 1 ...
Page 434: ...Appendix A Character Set A 2 DSA8300 Programmer Manual ...
Page 446: ...Appendix C Factory Default Setup Values C 8 DSA8300 Programmer Manual ...