Commands Listed in Alphabetical Order
Group
Acquisition
Syntax
ACQuire:STOPAfter:CONDition { ACQWfms | HISTWaveform |
HISTHits | MASKWaveform | MASKSamples | MASK<x>Hits |
MASKTOTalhit | FRAMecycle | AVGComp }
ACQuire:STOPAfter:CONDition?
Related Commands
Arguments
ACQWfms
sets the instrument to stop acquiring after some speci
fi
ed number of
raw acquisition cycles. This setting tells the instrument to count the number
of Main Timebase sweeps (Mag sweeps are not counted independently) and
stop acquisition after the speci
fi
ed number of acquisitions has been reached.
Use the ACQuire:STOPAfter:COUNt command to set the target number of
waveforms.
HISTWaveform
sets the instrument to stop acquisition after a
speci
fi
ed number of Histogram source waveforms is acquired. Use the
ACQuire:STOPAfter:COUNt command to set the speci
fi
ed number of
waveforms.
DSA8300 Programmer Manual
2-61
Summary of Contents for DSA8300
Page 1: ...xx DSA8300 Digital Serial Analyzer ZZZ Programmer Manual P077057002 077 0570 02 ...
Page 2: ......
Page 8: ...Preface iv DSA8300 Programmer Manual ...
Page 9: ...Getting Started ...
Page 10: ......
Page 15: ...Syntax and Commands ...
Page 16: ......
Page 404: ...Commands Listed in Alphabetical Order 2 388 DSA8300 Programmer Manual ...
Page 405: ...Status and Events ...
Page 406: ......
Page 426: ...Status and Events 3 20 DSA8300 Programmer Manual ...
Page 427: ...Examples ...
Page 428: ......
Page 431: ...Appendices ...
Page 432: ......
Page 433: ...Appendix A Character Set DSA8300 Programmer Manual A 1 ...
Page 434: ...Appendix A Character Set A 2 DSA8300 Programmer Manual ...
Page 446: ...Appendix C Factory Default Setup Values C 8 DSA8300 Programmer Manual ...