Commands Listed in Alphabetical Order
Examples
*CLS
clears the instrument status data structures.
COMPensate:ALLModules (No Query Form)
This command (no query form) compensates all installed module channels for
DC variances.
CAUTION.
Before sending this command, disconnect or disable all signals and
probes from all sampling input channels. Ensure that all electrical inputs are
terminated to 50
Ω
and all optical module inputs have dust covers installed before
continuing. For best results, ensure that the mainframe compensation is valid and
leave trigger sources for the selected channels connected and active.
NOTE.
Data from a successful compensation is not automatically saved into
nonvolatile user storage. Use the COMPensate:SAVe commands if you want to
retain the data in nonvolatile storage.
Group
Compensation
Syntax
COMPensate:ALLModules
Related Commands
COMPensate:SAVe:USER:ALLModules
Examples
COMPENSATE:ALLMODULES
performs the compensation routines on all installed
module channels.
COMPensate:CH<x> (No Query Form)
This command (no query form) compensates the module channel for DC
variances.
2-94
DSA8300 Programmer Manual
Summary of Contents for DSA8300
Page 1: ...xx DSA8300 Digital Serial Analyzer ZZZ Programmer Manual P077057002 077 0570 02 ...
Page 2: ......
Page 8: ...Preface iv DSA8300 Programmer Manual ...
Page 9: ...Getting Started ...
Page 10: ......
Page 15: ...Syntax and Commands ...
Page 16: ......
Page 404: ...Commands Listed in Alphabetical Order 2 388 DSA8300 Programmer Manual ...
Page 405: ...Status and Events ...
Page 406: ......
Page 426: ...Status and Events 3 20 DSA8300 Programmer Manual ...
Page 427: ...Examples ...
Page 428: ......
Page 431: ...Appendices ...
Page 432: ......
Page 433: ...Appendix A Character Set DSA8300 Programmer Manual A 1 ...
Page 434: ...Appendix A Character Set A 2 DSA8300 Programmer Manual ...
Page 446: ...Appendix C Factory Default Setup Values C 8 DSA8300 Programmer Manual ...