Commands Listed in Alphabetical Order
MEASUREMENT:MEAS2:REFLEVEL1:ABSOLUTE:MID?
might return
MEASUREMENT:MEAS2:REFLEVEL1:ABSOLUTE:MID 5.0000000000E–2
,
indicating that the mid reference level for Measurement 2 on Source 1 is set to
50 mV.
MEASUrement:MEAS<x>:REFLevel<x>:METHod
This command sets or queries the method the instrument uses to calculate
the reference levels for a speci
fi
ed measurement taken on a speci
fi
ed source
waveform. The measurement slot is speci
fi
ed by x, which can be 1 through 8. The
reference level is speci
fi
ed by x (which corresponds to the Source1 or Source2
waveform) and can be 1 or 2. Sending this command is equivalent to selecting
the option button for the type of calculation in the Reference Level Calc Method
section of the RefLevel dialog in the Measurement Setup dialog box.
The following diagram shows how RELative, ABSolute, HIDelta, and LODelta
methods calculate the High and Low reference levels.
Group
Measurement
DSA8300 Programmer Manual
2-241
Summary of Contents for DSA8300
Page 1: ...xx DSA8300 Digital Serial Analyzer ZZZ Programmer Manual P077057002 077 0570 02 ...
Page 2: ......
Page 8: ...Preface iv DSA8300 Programmer Manual ...
Page 9: ...Getting Started ...
Page 10: ......
Page 15: ...Syntax and Commands ...
Page 16: ......
Page 404: ...Commands Listed in Alphabetical Order 2 388 DSA8300 Programmer Manual ...
Page 405: ...Status and Events ...
Page 406: ......
Page 426: ...Status and Events 3 20 DSA8300 Programmer Manual ...
Page 427: ...Examples ...
Page 428: ......
Page 431: ...Appendices ...
Page 432: ......
Page 433: ...Appendix A Character Set DSA8300 Programmer Manual A 1 ...
Page 434: ...Appendix A Character Set A 2 DSA8300 Programmer Manual ...
Page 446: ...Appendix C Factory Default Setup Values C 8 DSA8300 Programmer Manual ...