![Tektronix DSA8300 Programmer'S Manual Download Page 298](http://html1.mh-extra.com/html/tektronix/dsa8300/dsa8300_programmers-manual_1077992298.webp)
Commands Listed in Alphabetical Order
The
<file path>
is a quoted string that de
fi
nes the
fi
le name and path. Input
the
fi
le path using the form <drive>\<dir>\<
fi
lename>. <drive> and one or more
<dir>s are optional. If you do not specify them, the instrument will read the
fi
le
from the default directory. The <
fi
lename> can be a Windows long
fi
le name. Do
not use wild card characters.
Examples
RECALL:SETUP FACTORY
recalls (and makes current) the instrument setup to its
factory defaults.
RECALL:SETUP "TEK00000.STP"
recalls the instrument setup from the
fi
le
TEK00000.STP in the default directory and on the default drive.
RECAll:WAVEform (No Query Form)
This command (no query form) recalls a stored waveform into a reference
location. Sending this command is equivalent to selecting Recall Waveform from
the File Menu, and then selecting a waveform
fi
le source and reference destination
in the Recall Waveform dialog box.
Group
Save and Recall
Syntax
RECAll:WAVEform <file path>,REF<x>
Arguments
<filepath>
is a quoted string that de
fi
nes the
fi
le name and path. Input the
fi
le path using the form <drive>/<dir>/<
fi
lename>. <drive> and one or more
<dir>s are optional. If you do not specify them, the instrument will recall the
waveform from the default directory. The <
fi
lename> can be a Windows long
fi
le name. Do not use wild card characters.
REF<x>
is the location in internal reference memory to which the waveform
is recalled.
Examples
RECALL:WAVEFORM "TEK00000.WFM",REF1
recalls the waveform stored in the
fi
le named TEK00000.WFM in the default directory and on the default drive to
Reference waveform 1.
REF<x>:POSition
This command allows you to set or query the vertical position of the speci
fi
ed
reference waveform. The reference waveform is speci
fi
ed by x, which can be
1 through 8. This is the equivalent to specifying a reference waveform in the
Waveform section of the Vertical Setup dialog box and then entering a value for
Position in the Setup section.
2-282
DSA8300 Programmer Manual
Summary of Contents for DSA8300
Page 1: ...xx DSA8300 Digital Serial Analyzer ZZZ Programmer Manual P077057002 077 0570 02 ...
Page 2: ......
Page 8: ...Preface iv DSA8300 Programmer Manual ...
Page 9: ...Getting Started ...
Page 10: ......
Page 15: ...Syntax and Commands ...
Page 16: ......
Page 404: ...Commands Listed in Alphabetical Order 2 388 DSA8300 Programmer Manual ...
Page 405: ...Status and Events ...
Page 406: ......
Page 426: ...Status and Events 3 20 DSA8300 Programmer Manual ...
Page 427: ...Examples ...
Page 428: ......
Page 431: ...Appendices ...
Page 432: ......
Page 433: ...Appendix A Character Set DSA8300 Programmer Manual A 1 ...
Page 434: ...Appendix A Character Set A 2 DSA8300 Programmer Manual ...