Commands Listed in Alphabetical Order
TRIGGER:CH2:CLKREC:RANGE?
might return
NONE
, indicating that continuous
clock recovery is not supported by the sampling module for Channel 2; therefore,
no custom user rate can be de
fi
ned for Channel 2.
TRIGger:CH<x>:CLKRec:USEr (No Query Form)
This command de
fi
nes the User (custom) clock rate (<nr3>) for the channel
speci
fi
ed (CH<x>), which can be 1 through 4. (The available choices are
dependent on the optical sampling module that is installed.) The channel
speci
fi
ed must be from an optical sampling module that provides continuous
clock-recovery capability. The clock rate de
fi
ned must be in the range available
for the channel/module chosen.
The query form returns the user-de
fi
ned rate in effect. The default rate is the
minimum value of the adjustable range.
Group
Trigger
Syntax
TRIGger:CH<x>:CLKRec:USEr <nr3>
Related Commands
SYSTem:PROPerties:CH<x>:CLKUser?
SYSTem:PROPerties:CH<x>:CLKRec?
Examples
TRIGGER:CH1:CLKREC:USER 10.52E9
de
fi
nes the user-speci
fi
ed clock
recovery rate as 10.52 Gb/s.
TRIGGER:CH1:CLKREC:USER?
might return
9.8E9
, indicating the current
user-de
fi
ned rate is 9.8 Gb/s.
TRIGger:CH<x>:CLKRec:VALue
This command sets or queries the clock recovery value that is applied to the
channel speci
fi
ed by x, which can be 1 through 4. The clock-recovery selections
require hardware support that may not be provided by your sampling module; see
your module user manual to determine which clock-recovery selections your
module supports. You can also use the SYStem:PROPerties:CH<x>:CLKRec?
query to determine clock recovery options available for the speci
fi
ed channel.
If you select a clock recovery setting that your sampling module does not support,
you will receive Execution Error code 220 with error message "Parameter Error."
(See Messages for general information on programming interface messages.)
DSA8300 Programmer Manual
2-321
Summary of Contents for DSA8300
Page 1: ...xx DSA8300 Digital Serial Analyzer ZZZ Programmer Manual P077057002 077 0570 02 ...
Page 2: ......
Page 8: ...Preface iv DSA8300 Programmer Manual ...
Page 9: ...Getting Started ...
Page 10: ......
Page 15: ...Syntax and Commands ...
Page 16: ......
Page 404: ...Commands Listed in Alphabetical Order 2 388 DSA8300 Programmer Manual ...
Page 405: ...Status and Events ...
Page 406: ......
Page 426: ...Status and Events 3 20 DSA8300 Programmer Manual ...
Page 427: ...Examples ...
Page 428: ......
Page 431: ...Appendices ...
Page 432: ......
Page 433: ...Appendix A Character Set DSA8300 Programmer Manual A 1 ...
Page 434: ...Appendix A Character Set A 2 DSA8300 Programmer Manual ...
Page 446: ...Appendix C Factory Default Setup Values C 8 DSA8300 Programmer Manual ...