Commands Listed in Alphabetical Order
Examples
MEASUREMENT:MEAS4:SOURCE1:GATE1:POS 1.9E–8
sets Gate 1 to 19 ns on
the Source 1 waveform used for Measurement 4 relative to the trigger.
MEASUREMENT:MEAS2:SOURCE1:GATE2:POS?
might return
MEASUREMENT:MEAS2:SOURCE1:GATE2:POS 2.29100000000E-008
,
indicating that Gate 2 is set at 22.91 ns on the Source 1 waveform used for
Measurement 2.
MEASUrement:MEAS<x>:SOUrce<x>:HILow? (Query Only)
This query only command returns all the High/Low settings for the speci
fi
ed
measurement taken on the speci
fi
ed source waveform. The measurement slot is
speci
fi
ed by x, which can be 1 through 8; the source is speci
fi
ed by x, which
can be 1 or 2.
NOTE.
The SOURCE<x> argument speci
fi
es the source, which can be source 1
or source 2. To set the actual waveform (channel, reference, or math) that is
assigned to source, you must use the MEASUrement:MEAS<x>:SOURCE<x>
WFM command. Two sources are provided because some measurements (gain,
phase) require two sources.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:SOUrce<x>:HILow?
Examples
MEASUREMENT:MEAS1:SOURCE1:HIGHLOW?
might return
MEASUREMENT:MEAS1:SOURCE1:HILOW:METHOD AUTO;
TRACK:HIGH:ENABLE 1;VALUE 2.00000000000E–001;
MEASUREMENT:MEAS1:SOURCE1:HILOW:TRACK:LOW:ENABLE 1;
VALUE 1.00000000000E–002
.
MEASUrement:MEAS<x>:SOUrce<x>:HILow:METHod
This command sets or queries the method of calculating the high/low levels for a
speci
fi
ed measurement on a speci
fi
ed source waveform. The measurement slot
is speci
fi
ed by x, which can be 1 through 8; the source is speci
fi
ed by x, which
can be 1 or 2. Sending this command is equivalent to specifying a measurement
slot (such as Meas 3), selecting a source (such as Source1), and then checking
the radio button for the desired Tracking Method in the in the Hi/Low pane of the
Measurement Setup dialog box.
The exact value of High and Low depends on which calculation method you set.
DSA8300 Programmer Manual
2-251
Summary of Contents for DSA8300
Page 1: ...xx DSA8300 Digital Serial Analyzer ZZZ Programmer Manual P077057002 077 0570 02 ...
Page 2: ......
Page 8: ...Preface iv DSA8300 Programmer Manual ...
Page 9: ...Getting Started ...
Page 10: ......
Page 15: ...Syntax and Commands ...
Page 16: ......
Page 404: ...Commands Listed in Alphabetical Order 2 388 DSA8300 Programmer Manual ...
Page 405: ...Status and Events ...
Page 406: ......
Page 426: ...Status and Events 3 20 DSA8300 Programmer Manual ...
Page 427: ...Examples ...
Page 428: ......
Page 431: ...Appendices ...
Page 432: ......
Page 433: ...Appendix A Character Set DSA8300 Programmer Manual A 1 ...
Page 434: ...Appendix A Character Set A 2 DSA8300 Programmer Manual ...
Page 446: ...Appendix C Factory Default Setup Values C 8 DSA8300 Programmer Manual ...