Commands Listed in Alphabetical Order
NOTE.
The SOURCE<x> argument speci
fi
es the source, which can be source 1
or source 2. To set the actual waveform (channel, reference, or math) that is
assigned to source, you must use the MEASUrement:MEAS<x>:SOURCE<x>
WFM command. Two sources are provided because some measurements (gain,
phase) require two sources.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:SOUrce<x>:HILow:TRACk:LOW:ENABle { ON
| OFF | NR1 }
MEASUrement:MEAS<x>:SOUrce<x>:HILow:TRACk:LOW:ENABle?
Related Commands
MEASUrement:MEAS<x>:SOUrce<x>:HILow:METHod
MEASUrement:MEAS<x>:SOUrce<x>:HILow:TRACk:HIGH:ENABle
Arguments
ON
enables tracking of the low value for the speci
fi
ed measurement on the
speci
fi
ed source waveform.
OFF
disables tracking of the low value for the speci
fi
ed measurement on the
speci
fi
ed source waveform.
NR1
set to 0 disables tracking of the low value for the speci
fi
ed measurement
on the speci
fi
ed source waveform; any other value enables tracking of the low
value for the speci
fi
ed measurement on the speci
fi
ed source waveform.
Examples
MEASUREMENT:MEAS2:SOURCE1:HILOW:TRACK:LOW:ENABLE ON
enables
tracking of the low value used to take Measurement 2 on the Source 1 waveform.
MEASUREMENT:MEAS3:SOURCE2:HILOW:TRACK:LOW:ENABLE?
might return
MEASUrement:MEAS3:SOURCE2:HILow:TRACK:LOW:ENABle 0, indicating
that the tracking of the low value for Measurement 3 on the Source 2 waveform is
disabled.
NOTE.
A query always returns a 0 or 1.
MEASUrement:MEAS<x>:SOUrce<x>:HILow:TRACk:LOW:VALue
This command sets or queries the low value used to calculate a speci
fi
ed
measurement on a speci
fi
ed source waveform. The measurement slot is speci
fi
ed
by x, which can be 1 through 8; the source is speci
fi
ed by x, which can be 1 or
2. Sending this command is equivalent to selecting a measurement slot (such
as Meas 3), selecting a source (such as Source1), setting tracking Low to off,
and then entering an absolute value in the Low control box in the in the Hi/Low
2-256
DSA8300 Programmer Manual
Summary of Contents for DSA8300
Page 1: ...xx DSA8300 Digital Serial Analyzer ZZZ Programmer Manual P077057002 077 0570 02 ...
Page 2: ......
Page 8: ...Preface iv DSA8300 Programmer Manual ...
Page 9: ...Getting Started ...
Page 10: ......
Page 15: ...Syntax and Commands ...
Page 16: ......
Page 404: ...Commands Listed in Alphabetical Order 2 388 DSA8300 Programmer Manual ...
Page 405: ...Status and Events ...
Page 406: ......
Page 426: ...Status and Events 3 20 DSA8300 Programmer Manual ...
Page 427: ...Examples ...
Page 428: ......
Page 431: ...Appendices ...
Page 432: ......
Page 433: ...Appendix A Character Set DSA8300 Programmer Manual A 1 ...
Page 434: ...Appendix A Character Set A 2 DSA8300 Programmer Manual ...
Page 446: ...Appendix C Factory Default Setup Values C 8 DSA8300 Programmer Manual ...