Commands Listed in Alphabetical Order
Measurement Setup dialog box. (If tracking is enabled for High, the instrument
automatically calculates the high value using the method set with the command,
MEASUrement:MEAS<x>:SOURCE<x>:HILow:METHod.)
NOTE.
The SOURCE<x> argument speci
fi
es the source, which can be source 1
or source 2. To set the actual waveform (channel, reference, or math) that is
assigned to source, you must use the MEASUrement:MEAS<x>:SOURCE<x>
WFM command. Two sources are provided because some measurements (gain,
phase) require two sources.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:SOUrce<x>:HILow:TRACk:HIGH:VALue <NR3
MEASUrement:MEAS<x>:SOUrce<x>:HILow:TRACk:HIGH:VALue?
Related Commands
MEASUrement:MEAS<x>:SOUrce<x>:HILow:METHod
MEASUrement:MEAS<x>:SOUrce<x>:HILow:TRACk:HIGH:ENABle
Arguments
NR3
is the absolute value of the high level for the HiLow settings.
Examples
MEASUREMENT:MEAS2:SOURCE1:HILOW:TRACK:HIGH:VALUE 5.0E-1
sets
the high value used to take Measurement 2 on the Source 1 waveform to 500 mV.
MEASUREMENT:MEAS1:SOURCE1:HILOW:TRACK:HIGH:VALUE?
might
return
MEASUREMENT:MEAS1:SOURCE1:HILOW:TRACK:HIGH VALUE
2.00000000000E-001
, indicating that the high value used to take Measurement
1on the Source 1 waveform is 200 mV.
MEASUrement:MEAS<x>:SOUrce<x>:HILow:TRACk:LOW:ENABle
This command sets or queries whether tracking of the low value for a speci
fi
ed
measurement on a speci
fi
ed source waveform is automatically done by the
instrument. The measurement slot is speci
fi
ed by x, which can be 1 through 8;
the source is speci
fi
ed by x, which can be 1 or 2. Sending this command is
equivalent to specifying a measurement slot (such as Meas 3), selecting a source
(such as Source1), and then checking the control box for Track Low in the in the
HiLow dialog of the Measurement Setup dialog box. (If tracking is enabled, the
instrument automatically calculates the low value using the method set with the
command, MEASUrement:MEAS<x>:SOURCE<x>:HILow:METHod.)
DSA8300 Programmer Manual
2-255
Summary of Contents for DSA8300
Page 1: ...xx DSA8300 Digital Serial Analyzer ZZZ Programmer Manual P077057002 077 0570 02 ...
Page 2: ......
Page 8: ...Preface iv DSA8300 Programmer Manual ...
Page 9: ...Getting Started ...
Page 10: ......
Page 15: ...Syntax and Commands ...
Page 16: ......
Page 404: ...Commands Listed in Alphabetical Order 2 388 DSA8300 Programmer Manual ...
Page 405: ...Status and Events ...
Page 406: ......
Page 426: ...Status and Events 3 20 DSA8300 Programmer Manual ...
Page 427: ...Examples ...
Page 428: ......
Page 431: ...Appendices ...
Page 432: ......
Page 433: ...Appendix A Character Set DSA8300 Programmer Manual A 1 ...
Page 434: ...Appendix A Character Set A 2 DSA8300 Programmer Manual ...
Page 446: ...Appendix C Factory Default Setup Values C 8 DSA8300 Programmer Manual ...