5-72
SYSTem:IP:GATeway<gate_adrs>(?)
Description
This command programs the gateway address for LAN operation. The programming must be performed
from either USB or GPIB controllers.
Parameters
Name
Range
Type
Description
<gate_adrs>
0 to 255
String
Programs the gateway address for LAN operation.
Programming must be performed from USB or GPIB
interfaces. Current gateway address can be observed
on LAN Properties front panel display.
Response
The 2074 will return the present IP address value similar to the following: 0.0.0.0
SYSTem:IP:HOSTname<name>(?)
Description
This command programs the host name address for LAN operation. The programming is performed in the
factory and it is highly suggested that users do not change the host name without first consulting a Tabor
customer service person.
Parameters
Name
Type
Description
<name>
String
Programs the host name for LAN operation.
Response
The 2074 will return a string containing the host name. String length is 16 characters.
SYSTem:KEEPalive:STATe{OFF|ON|0|1}(?)
Description
Use this command to toggle the keep alive mode on and off. The keep alive mode assures that LAN
connection remains uninterrupted throughout the duration of the LAN interfacing.
Parameters
Range
Type
Default
Description
0-1
Discrete
1
Toggles the keep alive mode on and off. When on, the
2074 constantly checks for smooth LAN connection at
intervals programmed by the syst:keep:time command.
The LAN will be probed as many times as programmed
by syst:keep:prob parameter to check if there is an
interruption
in
the
LAN
communication.
When
communication fails, the 2074 reverts automatically to
local (front panel) operation.
Response
The 2074 will return 0, or 1 depending on the present keep alive setting.
Summary of Contents for 5064
Page 26: ...1 5 Figure 1 2 ArbConnection The Control Panels Figure 1 3 ArbConnection The Wave Composer...
Page 71: ...2074 User Manual 2 18 This page intentionally left blank...
Page 145: ...2074 User Manual 3 74 Figure 3 48 Reading the 2074 Internal Temperature...
Page 220: ...4 ArbConnection Logging SCPI Commands 4 75 Figure 4 58 Log File Example...
Page 221: ...2074 User Manual 4 76 This page was intentionally left blank...
Page 333: ...2074 User Manual 7 2...
Page 359: ...2074 User Manual 7 28 This page was intentionally left blank...
Page 360: ...Appendices Appendix Title Page A Specifications A 1 1...
Page 361: ...2074 User Manual 2...