User Manual
6-8
Table
6-6, Square wave Characteristics
Parameter
Oscilloscope
Reading
Tested
Error
Limits CH1 CH2 CH3 CH4 Pass Fail
Rise/Fall
Time <4.5
ns
Ringing
<6 % + 10 mV
Over/undershoot
<6 % + 10 mV
Skew Between
Channels
Equipment: Oscilloscope, 50
Ω
, 20dB feedthrough attenuator
Preparation:
1. Configure the Oscilloscope follows:
Termination:
50
Ω
, 20 dB attenuator feed through at the
oscilloscope. Use identical cables to
connect the channels to the oscilloscope.
Setup:
As required for the test
2. Connect 2074 Channel outputs to the oscilloscope input
3. Configure the 2074 as follows:
Waveform: Arbitrary
SCLK: 200
MS/s
Amplitude: 6
V
Output: On
Test Procedure
1. Using ArbConnection prepare and download the following
waveform (all channels):
Wavelength: 1024
Waveform: Square
2. Measure the skew between channels 1 and 2 to be less than 2
ns.
3. Measure the skew between channels 3 and 4 to be less than 2
ns.
4. Measure the skew between channels 1 and 3 to be less than 7
ns.
5. Modify the phase offset parameters on channels 2, 3 and 4 from
-10 to 10 and check that the phase offset between channels is
increased by 5 ns with every phase offset step.
Test Results
Pass
Fail
Summary of Contents for 5064
Page 26: ...1 5 Figure 1 2 ArbConnection The Control Panels Figure 1 3 ArbConnection The Wave Composer...
Page 71: ...2074 User Manual 2 18 This page intentionally left blank...
Page 145: ...2074 User Manual 3 74 Figure 3 48 Reading the 2074 Internal Temperature...
Page 220: ...4 ArbConnection Logging SCPI Commands 4 75 Figure 4 58 Log File Example...
Page 221: ...2074 User Manual 4 76 This page was intentionally left blank...
Page 333: ...2074 User Manual 7 2...
Page 359: ...2074 User Manual 7 28 This page was intentionally left blank...
Page 360: ...Appendices Appendix Title Page A Specifications A 1 1...
Page 361: ...2074 User Manual 2...