7
Adjustments and Firmware Update
Channels 1, 2, 3 & 4 Adjustments
7-19
Terminate the 2074 output at the Oscilloscope input with the, 50
Ω
, 20 dB Feed through termination
3. Configure the 2074 as follows:
CAL:NUM 30, 66, 102 and 138 – respectively
Adjustment:
4. Adjust CAL:VAL to get the signal of 6 divisions on the screen.
CH1 – Setup 31
CH2 – Setup 67
CH3 – Setup 103
CH4 – Setup 139
50 MHz Amplitude, Amplifier Out
Equipment: 50
Ω
, 20 dB Feed through termination, Oscilloscope
Preparation:
1. Configure the Oscilloscope as follows:
Input
Impedance:
50
Ω
Range:
100
mV
2. Connect the 2074 Channel 1 output to the Oscilloscope input.
Terminate the 2074 output at the Oscilloscope input with the, 50
Ω
, 20 dB Feed through termination
3. Configure the 2074 as follows:
CAL:NUM 31, 67, 103 and 139 – respectively
Adjustment:
4. Adjust CAL:VAL to get the signal of 6 divisions on the screen.
CH1 – Setup 32
CH2 – Setup 68
CH3 – Setup 104
CH4 – Setup 140
50.00001 MHz Amplitude, Amplifier Out
Equipment: 50
Ω
, 20 dB Feed through termination, Oscilloscope
Preparation:
1. Configure the Oscilloscope as follows:
Input
Impedance:
50
Ω
Range:
100
mV
2. Connect the 2074 Channel 1 output to the Oscilloscope input.
Terminate the 2074 output at the Oscilloscope input with the, 50
Ω
, 20 dB Feed through termination
3. Configure the 2074 as follows:
CAL:NUM 32, 68, 104 and 140 – respectively
Adjustment:
4. Adjust CAL:VAL to get the signal of 6 divisions on the screen.
CH1 – Setup 33
CH2 – Setup 69
CH3 – Setup 105
CH4 – Setup 141
60 MHz Amplitude, Amplifier Out
Equipment: 50
Ω
, 20 dB Feed through termination, Oscilloscope
Preparation:
1. Configure the Oscilloscope as follows:
Input
Impedance:
50
Ω
Range:
100
mV
2. Connect the 2074 Channel 1 output to the Oscilloscope input.
Terminate the 2074 output at the Oscilloscope input with the, 50
Summary of Contents for 5064
Page 26: ...1 5 Figure 1 2 ArbConnection The Control Panels Figure 1 3 ArbConnection The Wave Composer...
Page 71: ...2074 User Manual 2 18 This page intentionally left blank...
Page 145: ...2074 User Manual 3 74 Figure 3 48 Reading the 2074 Internal Temperature...
Page 220: ...4 ArbConnection Logging SCPI Commands 4 75 Figure 4 58 Log File Example...
Page 221: ...2074 User Manual 4 76 This page was intentionally left blank...
Page 333: ...2074 User Manual 7 2...
Page 359: ...2074 User Manual 7 28 This page was intentionally left blank...
Page 360: ...Appendices Appendix Title Page A Specifications A 1 1...
Page 361: ...2074 User Manual 2...