Updating 2074
Firmware
WARNING
Only qualified persons may perform Firmware updates. DO
NOT even attempt to perform this operation unless you
were trained and certified by Tabor as you may inflict
damage on the instrument. Always verify with the factory
that you have the latest firmware file before you start with
your update.
Before you update the firmware of your 2074, check the revision level
which is installed in your instrument. Each firmware update was done
for a reason and therefore, if you want to update the firmware for a
problem in your system, check the readme file that is associated with
the update to see if an update will solve your problem. The revision
level of your firmware can be displayed as shown in Figure 7-3. To
access this screen, select the TOP menu, then select the Utility soft
key and scroll down to the System option. Press Enter and the screen
will show with the system information. Check both the Software
Version and the Version Date as both should match with the latest
release.
Figure
7-3, Software Version Screen
Summary of Contents for 5064
Page 26: ...1 5 Figure 1 2 ArbConnection The Control Panels Figure 1 3 ArbConnection The Wave Composer...
Page 71: ...2074 User Manual 2 18 This page intentionally left blank...
Page 145: ...2074 User Manual 3 74 Figure 3 48 Reading the 2074 Internal Temperature...
Page 220: ...4 ArbConnection Logging SCPI Commands 4 75 Figure 4 58 Log File Example...
Page 221: ...2074 User Manual 4 76 This page was intentionally left blank...
Page 333: ...2074 User Manual 7 2...
Page 359: ...2074 User Manual 7 28 This page was intentionally left blank...
Page 360: ...Appendices Appendix Title Page A Specifications A 1 1...
Page 361: ...2074 User Manual 2...