5-78
Service Request
Enable Register
(SRE)
The Service Request enable register is an 8-bit register that
enables corresponding summary messages in the Status Byte
Register. Thus, the application programmer can select reasons for
the generator to issue a service request by altering the contents of
the Service Request Enable Register.
The Service Request Enable Register is read with the *SRE?
common query. The response to this query is a number that
represents the sum of the binary-weighted value of the Service
Request Enable Register. The value of the unused bit 6 is always
zero.
The Service Request Enable Register is written using the *SRE
command followed by a decimal value representing the bit values of
the Register. A bit value of 1 indicates an enabled condition.
Consequently, a bit value of zero indicates a disabled condition.
The Service Request Enable Register is cleared by sending *SRE0.
The generator always ignores the value of bit 6. Summary of *SRE
commands is given in the following.
*SRE0
- Clears all bits in the register.
*SRE1
- Not used.
*SRE2
- Not used.
*SRE4
- Not used.
*SRE8
- Not used.
*SRE16
- Service request on MAV.
*SRE32
- Service request on ESB summary bit.
*SRE128
- Not used.
Standard Event
Status Register
(ESR)
The Standard Event Status Register reports status for special
applications. The 8 bits of the ESR have been defined by the IEEE-
STD-488.2 as specific conditions, which can be monitored and
reported back to the user upon request. The Standard Event Status
Register is destructively read with the *ESR? common query. The
Standard Event Status Register is cleared with a *CLS common
command, with a power-on and when read by *ESR?.
The arrangement of the various bits within the register is firm and is
required by all GPIB instruments that implement the IEEE-STD-
488.2. Description of the various bits is given in the following:
Bit 0
- Operation Complete. Generated in response to the *OPC
command. It indicates that the device has completed all selected
and pending operations and is ready for a new command.
Bit 1
- Request Control. This bit operation is disabled on the Model
2074.
Bit 2
- Query Error. This bit indicates that an attempt is being made
to read data from the output queue when no output is either present
Summary of Contents for 5064
Page 26: ...1 5 Figure 1 2 ArbConnection The Control Panels Figure 1 3 ArbConnection The Wave Composer...
Page 71: ...2074 User Manual 2 18 This page intentionally left blank...
Page 145: ...2074 User Manual 3 74 Figure 3 48 Reading the 2074 Internal Temperature...
Page 220: ...4 ArbConnection Logging SCPI Commands 4 75 Figure 4 58 Log File Example...
Page 221: ...2074 User Manual 4 76 This page was intentionally left blank...
Page 333: ...2074 User Manual 7 2...
Page 359: ...2074 User Manual 7 28 This page was intentionally left blank...
Page 360: ...Appendices Appendix Title Page A Specifications A 1 1...
Page 361: ...2074 User Manual 2...