2074
User Manual
6-24
2.
Connect 2074 CH1 output to the distortion analyzer input.
3.
Configure the 2074 as follows:
SCLK: 200
MS/s
Waveform: Arbitrary
Output: On
4. Using ArbConnection prepare and download the following
waveform:
Wavelength:
1 M points (2 M with an option installed)
Waveform: Sine
wave
Test Procedure
1.
Check the resulting trace on the oscilloscope
2.
Perform Sine wave distortion. It should be less than 0.1 %
Test Results
Pass
Fail
USB Control
Equipment: Distortion Analyzer, ArbConnection
Preparation:
1.
Set up the 2074 for USB operation and connect the instrument
to a host controller
2.
Connect 2074 CH1 output to the distortion analyzer input.
3.
Configure the 2074 as follows:
SCLK: 200
MS/s
Waveform: Arbitrary
Output: On
4. Using ArbConnection prepare and download the following
waveform:
Wavelength:
1 M points (2 M with an option installed)
Waveform: Sine
wave
Test Procedure
1.
Check the resulting trace on the oscilloscope
2.
Perform Sine wave distortion. It should be less than 0. 1 %
Test Results
Pass
Fail
LAN Control
Equipment: Distortion Analyzer, ArbConnection
Preparation:
1.
Set up the 2074 for LAN operation and connect the
instrument to a host controller
2.
Connect 2074 CH1 output to the distortion analyzer input.
3.
Configure the 2074 as follows:
SCLK: 200
MS/s
Summary of Contents for 5064
Page 26: ...1 5 Figure 1 2 ArbConnection The Control Panels Figure 1 3 ArbConnection The Wave Composer...
Page 71: ...2074 User Manual 2 18 This page intentionally left blank...
Page 145: ...2074 User Manual 3 74 Figure 3 48 Reading the 2074 Internal Temperature...
Page 220: ...4 ArbConnection Logging SCPI Commands 4 75 Figure 4 58 Log File Example...
Page 221: ...2074 User Manual 4 76 This page was intentionally left blank...
Page 333: ...2074 User Manual 7 2...
Page 359: ...2074 User Manual 7 28 This page was intentionally left blank...
Page 360: ...Appendices Appendix Title Page A Specifications A 1 1...
Page 361: ...2074 User Manual 2...