6
Performance Checks
Test Procedures
6-15
Frequency: 1
MHz
Waveform: Sine
wave
Run Mode:
Triggered
Output: On
Test Procedure
1. Toggle 2074 trigger slope from positive to negative visa versa
2. Verify on the oscilloscope that the 2074 transitions are
synchronized with the slope of the trigger
Test Results
Pass
Fail
Trigger Level
Equipment: Oscilloscope, function generator
Preparation:
1. Configure the Oscilloscope as follows:
Termination:
20 dB, 50
Ω
feedthrough attenuator at the
oscilloscope input
Setup:
As required for the test
2. Connect 2074 Channel 1 output to the oscilloscope input
3. Configure the function generator as follows:
Frequency 10
kHz
Run Mode:
Continuous
Waveform: Squarewave.
Amplitude:
1 V
4. Connect the function generator output to the 2074 TRIG IN
connector
5. Configure the 2074 as follows:
Frequency: 1
MHz
Waveform: Sine
wave
Run Mode:
Triggered
Trigger level:
0 V
Ch1 Output:
On
Test Procedure
1. Verify that the 2074 outputs triggered waveforms spaced at 0.1
ms
2. Modify the function generator offset to +2 V and change the
2074 trigger level to +4 V. Verify that the 2074 outputs triggered
waveforms spaced at 0.1 ms
3. Modify the function generator offset to -2 V and change the
2074 trigger level to -4 V. Verify that the 2074 outputs triggered
waveforms spaced at 0.1 ms
Test Results
Pass
Fail
Summary of Contents for 5064
Page 26: ...1 5 Figure 1 2 ArbConnection The Control Panels Figure 1 3 ArbConnection The Wave Composer...
Page 71: ...2074 User Manual 2 18 This page intentionally left blank...
Page 145: ...2074 User Manual 3 74 Figure 3 48 Reading the 2074 Internal Temperature...
Page 220: ...4 ArbConnection Logging SCPI Commands 4 75 Figure 4 58 Log File Example...
Page 221: ...2074 User Manual 4 76 This page was intentionally left blank...
Page 333: ...2074 User Manual 7 2...
Page 359: ...2074 User Manual 7 28 This page was intentionally left blank...
Page 360: ...Appendices Appendix Title Page A Specifications A 1 1...
Page 361: ...2074 User Manual 2...