1. Each channel has its own segment table buffer. Therefore, make sure you selected the correct active
channel (with the INST:SEL command) before you download segment table data to the generator
2. Minimum number of segments is 1; maximum number of segments is 16k
3. Maximum segment size depends on your installed option. With the basic 2074 you can program
maximum 1M in one segment. With the 2M option, you can use the full size of 2 Meg
4. Segment table data has 32-bit values of which are used for segment size. Therefore, Data for each
segment must have 4 bytes
5. The number of bytes in a complete segment table must divide by 6. The Model 2074 has no control
over data sent to its segment table during data transfer. Therefore, wrong data and/or incorrect number
of bytes will cause erroneous memory partition
Parameters
Name
Type
Description
<binary_block>
Binary
Block of binary data that contains information on the
segment table.
Summary of Contents for 5064
Page 26: ...1 5 Figure 1 2 ArbConnection The Control Panels Figure 1 3 ArbConnection The Wave Composer...
Page 71: ...2074 User Manual 2 18 This page intentionally left blank...
Page 145: ...2074 User Manual 3 74 Figure 3 48 Reading the 2074 Internal Temperature...
Page 220: ...4 ArbConnection Logging SCPI Commands 4 75 Figure 4 58 Log File Example...
Page 221: ...2074 User Manual 4 76 This page was intentionally left blank...
Page 333: ...2074 User Manual 7 2...
Page 359: ...2074 User Manual 7 28 This page was intentionally left blank...
Page 360: ...Appendices Appendix Title Page A Specifications A 1 1...
Page 361: ...2074 User Manual 2...