E
r
r
o
r
!
B
o
o
k
m
a
r
k
n
o
t
5-61
Parameters
Name
Range
Type
Default
Description
<high>
-4.990 to 5
Numeric
5
Will set the pulse high level in units of volts. Note that
the high level setting must be higher than the low
level setting. Also note that high to low level value
must be equal or larger than 10 mV.
Response
The 2074 will return the present low level value in unit of volts.
AUXiliary:PULse:LEVel:LOW<low>(?)
Description
This command will program the phase offset between two adjacent instruments. Normally this command
should be used on the slave unit. The phase offset control provides means of generating multiple signals
with phase offset between them.
Parameters
Name
Range
Type
Default
Description
<low>
-5 to 4.990
Numeric
0
Will set the pulse low level in units of volts. Note that
the low level setting must be smaller than the high
level setting. Also note that low to high level value
must be equal or larger than 10 mV.
Response
The 2074 will return the present high level value in unit of volts.
AUXiliary:PULse:PERiod<period>(?)
Description
This command will program the pulse repetition rate (period). Note that the sum of all parameters, including
the pulse delay, rise, high and fall times must not exceed the programmed pulse period and therefore, it is
recommended that the pulse period be programmed first before all other pulse parameters. Note that by
selecting the double pulse mode, the pulse period remains unchanged.
Parameters
Name
Range
Type
Default
Description
<period>
80e-9 to 1e6
Numeric
10e-3
Will program the period of the pulse waveform in
units of seconds. The maximum period is extended to
2e6 when the 2 Meg memory option is installed.
Response
The 2074 will return the present pulse period value in units of seconds.
Summary of Contents for 5064
Page 26: ...1 5 Figure 1 2 ArbConnection The Control Panels Figure 1 3 ArbConnection The Wave Composer...
Page 71: ...2074 User Manual 2 18 This page intentionally left blank...
Page 145: ...2074 User Manual 3 74 Figure 3 48 Reading the 2074 Internal Temperature...
Page 220: ...4 ArbConnection Logging SCPI Commands 4 75 Figure 4 58 Log File Example...
Page 221: ...2074 User Manual 4 76 This page was intentionally left blank...
Page 333: ...2074 User Manual 7 2...
Page 359: ...2074 User Manual 7 28 This page was intentionally left blank...
Page 360: ...Appendices Appendix Title Page A Specifications A 1 1...
Page 361: ...2074 User Manual 2...