6-28
3. Change the function generator run mode to Burst and set Burst
Count to 100
4. Press the Reset/Arm button on the 2074 to reset and arm the
totalize function
5. Manually trigger the function generator and verify that the 2074
counter reading is 100
±
1
Test Results
Pass
Fail
Totalize, Infinite
Equipment: Function Generator with at least 1 ppm accuracy
Preparation:
1. Configure the function generator as follows:
Frequency: 150
MHz
Wave:
Square
Amplitude 500
mV
2. Connect the function generator to the 2074 TRIG IN connector
3. Configure the 2074, as follows:
Auxiliary Function: Counter/Timer
Function: Totalize,
Infinite
Trigger Level:
0 V
Test Procedure:
1. Verify that the display is updated continuously with the totalized
counts
2. Verify that the reading is held at 4000000000 counts and that
the overflow indication turns on
Test Results
Pass
Fail
3. Change the function generator run mode to Burst and set Burst
Count to 1’000’000
4. Press the Reset/Arm button on the 2074 to reset and arm the
totalize function
5. Manually trigger the function generator and verify that the 2074
counter reading is 1000000
±
2
Test Results
Pass
Fail
Summary of Contents for 5064
Page 26: ...1 5 Figure 1 2 ArbConnection The Control Panels Figure 1 3 ArbConnection The Wave Composer...
Page 71: ...2074 User Manual 2 18 This page intentionally left blank...
Page 145: ...2074 User Manual 3 74 Figure 3 48 Reading the 2074 Internal Temperature...
Page 220: ...4 ArbConnection Logging SCPI Commands 4 75 Figure 4 58 Log File Example...
Page 221: ...2074 User Manual 4 76 This page was intentionally left blank...
Page 333: ...2074 User Manual 7 2...
Page 359: ...2074 User Manual 7 28 This page was intentionally left blank...
Page 360: ...Appendices Appendix Title Page A Specifications A 1 1...
Page 361: ...2074 User Manual 2...