2-16
764333-675
The following displays with CALIB flashing:
CALIB RESTORE
VIEW PRINT
Select
CALIB
and press the
SAVE VALUES
pushbutton.
Select
MORE
and press the
SAVE VALUES
touch pad.
Select
DEFAULT
and press the
SAVE VALUES
touch
pad to load default calibration values.
• Restore Calibration Data
Select
CALIB
, then press
SAVE VALUES
touch pad.
Display shows:
CALIB RESTORE
VIEW PRINT
Select
RESTORE
, then press
SAVE VALUES
touch
pad. Display shows:
PRES TEMP
CONC UV
The routines are similar for PRESS, TEMP, and CONC.
When a selection is made, the following is displayed:
WARNING! CALIB
WILL BE CHANGED
Press
SAVE VALUES
touch pad. Display shows data
points that can be changed, such as LOW RAW, HIGH
RAW, LOW PT, HIGH PT. Temperatures use all points,
CONC uses only LOW and HIGH RAW, and
PRESSURE uses LOW RAW only. Data can be entered
using the keypad. If the previous calibration data
needs to be re-entered due to calibration error, use the
print function described later in this section
(PRINT)
.
Press
SAVE VALUES
touch pad. Display shows:
* UPDATING *
* MEMORY *
Display shows:
PRES TEMP
CONC MORE
Select
CANCEL
to continue. Display shows:
MFG TESTS
VALUES CALIB
• View
Current real time value of pressure, temperature,
concentration and UV can be viewed in this routine.
Select
CALIB
, then press
SAVE VALUES
touch pad.
Select
VIEW,
then press
SAVE VALUES
touch pad.
Display shows:
PRES TEMP
CONC UV
Selecting
PRESS/TEMP/CONC/UV
now displays the
current, real-time reading. Press
CANCEL
after
reading an item. Re-enter VIEW if other points need to
be read.
Selection of this function allows all current calibration
data to be printed. See Figure 2-9 for an example.
Summary of Contents for SYSTEM 1E
Page 1: ...MAINTENANCE MANUAL SYSTEM 1E Liquid Chemical Sterilant Processing System 12 09 10 P764333 674 ...
Page 3: ......
Page 71: ...5 19 764333 675 Figure 5 2 Analog Cable Assembly Pin Out Locations ...
Page 74: ...5 22 764333 675 Figure 5 7 DC Power Supply Harness Pin Out Locations PS P1 ...
Page 96: ...7 19 764333 675 Figure 7 17 Aquafine STERIUV5 Control Panel The Control Panel A B 3 2 1 ...
Page 108: ......
Page 110: ...ii 764333 676 Example 1 How To Use IPB Page Typical ...
Page 119: ...10 764333 676 Figure 1 SYSTEM 1E Piping Schematic ...
Page 121: ...12 764333 676 Figure 2 SYSTEM 1E Complete 4 3 2 ...
Page 128: ...19 764333 676 Figure 5 Fluid and Air Components 2 of 2 51 53 48 47 50 49 52 ...
Page 133: ...24 764333 676 Figure 7 Float Block Switch Assembly 1 5 8 7 5 4 10 4 2 6 9 9 3 CK2 CK3 LS3 CK1 ...
Page 137: ...28 764333 676 Figure 9 Drain Block Assembly 3 4 5 6 7 3 1 2 8 CK8 TC3 ...
Page 143: ...34 764333 676 Figure 12 Bulkhead Connector Assembly 1 2 2 2 3 1 1 3 5 4 4 6 ...
Page 155: ...46 764333 676 Figure 18 Harness Cable Assemblies 1 of 2 1 2 3 4 5 6 P10 PT1 P5B P11 P2 P5A ...
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......