2-17
764333-675
2.2 ENHANCED OPERATOR MODE
When this mode is selected, display reverts to a
standard operator mode screen; however, while cycle
is progressing, pressing
STATUS
button allows
display to show various parameters that are not seen
in normal operation. These additional displays are as
follows:
• Thermocouple temperatures TC1, TC2 and TC3
(read in degrees C)
• Concentration monitor
• Pressure (read by transducer, read in PSI)
• UV Intensity (read by UV monitor)
• Cycle phase
* Time = length of time in phase
** Conc = Concentration count at end of phase
Under normal conditions, display shows
OPERATING, with a projected time to completion.
Pressing the
STATUS
button shows the first alternate
display. Pressing again shows second alternate
display, and so on. Pressing
PAPER FEED
touch pad
forces unit to print current status of LS6 and float
switch.
NOTE: The display reverts back to the OPERATING
display after a few seconds, so it is necessary to
continue pressing the STATUS button to continue
showing the alternate displays.
INLET TEMPERATURE
LOW PT =XXXX
LOW RAW =XXXX
HIGH PT =XXXX
HIGH RAW =XXXXX
OUTLET TEMPERATURE
LOW PT =XXXX
LOW RAW =XXXX
HIGH PT =XXXX
HIGH RAW =XXXXX
DRAIN TEMPERATURE
LOW PT =XXXX LOW RAW =XXXX
HIGH PT =XXXX
HIGH RAW =XXXXX
CONCENTRATION MONITOR
LOW RAW = XXXX
HIGH RAW = XXXX
PRESSURE TRANSDUCER
LOW RAW = XXXX
UV LAMP SENSOR
LOW PT = X.X mA
LOW RAW = XXXXX
HIGH PT = XX.X mA
HIGH RAW = XXXXX
Figure 2-9. Calibration Data Printout
SYSTEM 1E
DATE: mm-dd-yy
CYCLE START: hh:mm:ssp
OPERATOR:_______________
LOAD ID:________________
REMARKS:________________
________________
________________
________________
==========================
TIME* CONC**
WARM/MIX XX.X MIN XXX
EXPOSURE XX.X MIN XXX
RINSE: 1 XX.X SEC XXX
RINSE: 2 XX.X SEC XXX
DRAIN: XX.X SEC XXX
TOTAL: XX.X MIN XXX
==========================
TEMP: XX.X - XX.X DEG. C
CONCENTRATION: XXX
EXPOSURE TIME: X.X
FILL TIME: X.X
INLET TEMP: XX DEG. C
OPERATOR: ______________
SERIAL # XXXXXXXXXX
CYCLE COUNT: XXX
CYCLE COMPLETE:hh:mm:ssp
CHAMBER OPENED:hh:mm:ssp
Figure 2-10. Cycle Data
Printout
Summary of Contents for SYSTEM 1E
Page 1: ...MAINTENANCE MANUAL SYSTEM 1E Liquid Chemical Sterilant Processing System 12 09 10 P764333 674 ...
Page 3: ......
Page 71: ...5 19 764333 675 Figure 5 2 Analog Cable Assembly Pin Out Locations ...
Page 74: ...5 22 764333 675 Figure 5 7 DC Power Supply Harness Pin Out Locations PS P1 ...
Page 96: ...7 19 764333 675 Figure 7 17 Aquafine STERIUV5 Control Panel The Control Panel A B 3 2 1 ...
Page 108: ......
Page 110: ...ii 764333 676 Example 1 How To Use IPB Page Typical ...
Page 119: ...10 764333 676 Figure 1 SYSTEM 1E Piping Schematic ...
Page 121: ...12 764333 676 Figure 2 SYSTEM 1E Complete 4 3 2 ...
Page 128: ...19 764333 676 Figure 5 Fluid and Air Components 2 of 2 51 53 48 47 50 49 52 ...
Page 133: ...24 764333 676 Figure 7 Float Block Switch Assembly 1 5 8 7 5 4 10 4 2 6 9 9 3 CK2 CK3 LS3 CK1 ...
Page 137: ...28 764333 676 Figure 9 Drain Block Assembly 3 4 5 6 7 3 1 2 8 CK8 TC3 ...
Page 143: ...34 764333 676 Figure 12 Bulkhead Connector Assembly 1 2 2 2 3 1 1 3 5 4 4 6 ...
Page 155: ...46 764333 676 Figure 18 Harness Cable Assemblies 1 of 2 1 2 3 4 5 6 P10 PT1 P5B P11 P2 P5A ...
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......