5.2.8
Thermocouple Failed
Description
– Control monitors TC1 and TC2 and
alarm will occur if readings are more than 3°C apart.
Action:
1. Loose connection to TC1 or TC2 – repair
connection. (Refer to Figure 5-2.)
2. Faulty TC1 or TC2 – replace.
3. Out of Calibration – recalibrate.
4. Faulty interface board – replace.
5. Faulty control board – replace.
6. Tray drain strainer obstructed – clean or replace.
7. Faulty P1 – replace.
8. Faulty SSR4 – replace. (Refer to Figure 5-5.)
5.2.9
Circulation Valve Fault
Description
– Control monitors circulation pressure
(CP) and pressure with circulation pump (P1) on and
circulation valve closed (CZ). Alarm occurs if CP–CZ
< 0.975 psi (2.5 counts).
Action:
1. V5 leaking – replace.
2. Faulty SOL3 – replace. (Refer to Figure 5-5.)
3. Incomplete tray being run (no container or lid) –
retest with complete tray.
4. PT1 fault – repair. (Refer to Figure 5-1 and 5-2.)
5. Faulty CK11 – replace.
5.2.10 Heat Rate Low
Description
– Control monitors temperature rise of
fluid during heating phase. Alarm occurs if
temperature does not rise an average of 1.0°C per
minute.
Action:
1. Faulty HTR1 – replace.
2. LS1 tripped – reset.
3. Faulty SSR1 – replace. (Refer to Figure 5-5.)
4. Faulty control board – replace.
5.2.11 Drain Valve Fault
Description
– Control monitors LS3 just prior to
draining the chamber. If LS3 switch does not sense a
full condition, alarm occurs.
Action:
1. Chamber leak through cracked tray – replace tray.
2. Faulty V5 – replace.
3. Faulty LS3 – replace.
4. Faulty CK8 – rebuild.
5. CK2 or CK3 leaking – replace.
6. Plumbing component leak – repair.
5.2.12 Drain Time Fault
Description
– Control monitors the LS3 after the unit
starts draining. If LS3 does not sense a not-full
condition within five seconds, alarm occurs.
Action:
1. Clogged FLT2 – replace.
2. Faulty CK1 – replace.
3. Drain hose restricted – repair.
4. Faulty V5 – replace.
5. Faulty LS3 – replace.
5.2 DIAGNOSTIC CYCLE FAULTS (CONTINUED)
Summary of Contents for SYSTEM 1E
Page 1: ...MAINTENANCE MANUAL SYSTEM 1E Liquid Chemical Sterilant Processing System 12 09 10 P764333 674 ...
Page 3: ......
Page 71: ...5 19 764333 675 Figure 5 2 Analog Cable Assembly Pin Out Locations ...
Page 74: ...5 22 764333 675 Figure 5 7 DC Power Supply Harness Pin Out Locations PS P1 ...
Page 96: ...7 19 764333 675 Figure 7 17 Aquafine STERIUV5 Control Panel The Control Panel A B 3 2 1 ...
Page 108: ......
Page 110: ...ii 764333 676 Example 1 How To Use IPB Page Typical ...
Page 119: ...10 764333 676 Figure 1 SYSTEM 1E Piping Schematic ...
Page 121: ...12 764333 676 Figure 2 SYSTEM 1E Complete 4 3 2 ...
Page 128: ...19 764333 676 Figure 5 Fluid and Air Components 2 of 2 51 53 48 47 50 49 52 ...
Page 133: ...24 764333 676 Figure 7 Float Block Switch Assembly 1 5 8 7 5 4 10 4 2 6 9 9 3 CK2 CK3 LS3 CK1 ...
Page 137: ...28 764333 676 Figure 9 Drain Block Assembly 3 4 5 6 7 3 1 2 8 CK8 TC3 ...
Page 143: ...34 764333 676 Figure 12 Bulkhead Connector Assembly 1 2 2 2 3 1 1 3 5 4 4 6 ...
Page 155: ...46 764333 676 Figure 18 Harness Cable Assemblies 1 of 2 1 2 3 4 5 6 P10 PT1 P5B P11 P2 P5A ...
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......