5-10
764333-675
5.2.5
Fill time > 5 Minutes
Description
– Control monitors LS3 from start of fill.
If full condition is not sensed within five minutes,
alarm occurs.
Action:
1. Water pressure too low – correct (refer to
S
ECTION
1
for specifications).
2. MaxPure Filter may be clogged – replace.
3. If MaxPure Filter replacement does not correct
problem – replace A prefilter, then B prefilter if
problem persists.
4. Restriction in drain hose – correct.
5. Faulty SOL7 fill valve – replace. (Refer to Figure 5-5.)
6. Faulty CK4/strainer – repair or replace.
7. Faulty CK2 and or CK3 – replace.
8. Faulty V5 – repair.
9. Faulty SOL4 – replace. (Refer to Figure 5-5.)
10.Faulty SOL5 – replace. (Refer to Figure 5-5.)
11.Faulty LS3 – replace. (Refer to Figure 5-3.)
12.Faulty LS2– replace.
5.2.6
Pressure Transducer Fault
Description
– Control monitors fill pressure (FP) and
fill zero (FZ). If FP < 5.85 psi or if FZ is > 1.95 psi, alarm
occurs.
Action:
If printout shows inlet < 5.85
1. Water pressure too low – correct.
2. Filters plugged – replace.
3. Faulty SOL7 fill valve – replace.
4. Faulty CK4/strainer – repair or replace.
5. Faulty PT1 – recalibrate or replace. (Refer to
6. Float block not drained from previous cycle – drain
and re-test.
7. Faulty LS3 – replace. (Refer to Figure 5-3.)
If printout shows zero > 1.95
1. Faulty PT1 – recalibrate or replace. (Refer to
2. Faulty CK2 and or CK3 – replace.
3. Faulty MaxPure™ Filter – replace.
5.2.7
Circulation Pressure Low
Description
– Control monitors circulation pressure
(CP) and fill zero (FZ) and performs a calculation. If
CP < (1.073 psi + FZ), alarm occurs.
Action:
1. Incomplete tray being run (no container or lid) –
retest with complete tray.
2. S40 being run – run without S40.
3. Drain screen plugged – clean.
4. Faulty CK2 and/or CK3 – replace.
5. Faulty P1 – replace.
6. Faulty SSR4 – replace. (Refer to Figure 5-5.)
7. PT1 fault – repair. (Refer to Figure 5-1 and 5-2.)
8. V6 – repair.
9. Faulty CK11 – replace.
5.2 DIAGNOSTIC CYCLE FAULTS (CONTINUED)
Summary of Contents for SYSTEM 1E
Page 1: ...MAINTENANCE MANUAL SYSTEM 1E Liquid Chemical Sterilant Processing System 12 09 10 P764333 674 ...
Page 3: ......
Page 71: ...5 19 764333 675 Figure 5 2 Analog Cable Assembly Pin Out Locations ...
Page 74: ...5 22 764333 675 Figure 5 7 DC Power Supply Harness Pin Out Locations PS P1 ...
Page 96: ...7 19 764333 675 Figure 7 17 Aquafine STERIUV5 Control Panel The Control Panel A B 3 2 1 ...
Page 108: ......
Page 110: ...ii 764333 676 Example 1 How To Use IPB Page Typical ...
Page 119: ...10 764333 676 Figure 1 SYSTEM 1E Piping Schematic ...
Page 121: ...12 764333 676 Figure 2 SYSTEM 1E Complete 4 3 2 ...
Page 128: ...19 764333 676 Figure 5 Fluid and Air Components 2 of 2 51 53 48 47 50 49 52 ...
Page 133: ...24 764333 676 Figure 7 Float Block Switch Assembly 1 5 8 7 5 4 10 4 2 6 9 9 3 CK2 CK3 LS3 CK1 ...
Page 137: ...28 764333 676 Figure 9 Drain Block Assembly 3 4 5 6 7 3 1 2 8 CK8 TC3 ...
Page 143: ...34 764333 676 Figure 12 Bulkhead Connector Assembly 1 2 2 2 3 1 1 3 5 4 4 6 ...
Page 155: ...46 764333 676 Figure 18 Harness Cable Assemblies 1 of 2 1 2 3 4 5 6 P10 PT1 P5B P11 P2 P5A ...
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......