2-15
764333-675
The following displays and prints:
SET UV TO LO CUR
PRESS SV
NOTE: "Low and High Raw" values may be different.
Set signal generator to 4.0 mA.
Press the
SAVE VALUES
pushbutton.
The following displays for 30 seconds:
STABILIZING LOW
UV = 4.0 mA
The following displays with the first number of the UV
intensity flashing:
LO SET PT 04.3
PRESS SV TO SET
Set to 4.0 and press the
SAVE VALUES
pushbutton.
The following displays:
SET UV TO HI CUR
PRESS SV
Press the
SAVE VALUES
pushbutton.
The following displays:
STABILIZING HIGH
UV = 20.0 mA
While the
STABILIZING HIGH
screen is shown
quickly raise signal generator to 20.0 mA. Wait 30
seconds from the time the value stopped changing.
The following displays with the first number of the UV
intensity flashing:
HI SET PT 20.0
PRESS SV TO SET
Set to 20.0 and press the
SAVE VALUES
pushbutton.
The following displays for a few seconds:
* UPDATING *
* MEMORY *
Then the follow displays and prints:
* UV SENSOR *
* CALIBRATED *
NOTE: "Low and High Raw" values may be different.
Press the
SAVE VALUES
pushbutton.
The following displays with CALIB flashing:
MFG TESTS
VALUES CALIB
• Default Calibration
Select
CALIB
and press the
SAVE VALUES
pushbutton.
========================
===== UV INTENSITY =====
===== CALIBRATION ======
========================
AT 12:33:32P
ON 8-07-08
PREVIOUS VALUES
UV LAMP SENSOR
------------------------
LOW PT = X.X
LOW RAW = XXXXX
HIGH PT = XX.X
HIGH RAW = XXXXX
CALIBRATED VALUES
UV LAMP SENSOR
------------------------
LOW PT = X.X mA
LOW RAW = XXXXX
HIGH PT = XX.X mA
HIGH RAW = XXXXX
Summary of Contents for SYSTEM 1E
Page 1: ...MAINTENANCE MANUAL SYSTEM 1E Liquid Chemical Sterilant Processing System 12 09 10 P764333 674 ...
Page 3: ......
Page 71: ...5 19 764333 675 Figure 5 2 Analog Cable Assembly Pin Out Locations ...
Page 74: ...5 22 764333 675 Figure 5 7 DC Power Supply Harness Pin Out Locations PS P1 ...
Page 96: ...7 19 764333 675 Figure 7 17 Aquafine STERIUV5 Control Panel The Control Panel A B 3 2 1 ...
Page 108: ......
Page 110: ...ii 764333 676 Example 1 How To Use IPB Page Typical ...
Page 119: ...10 764333 676 Figure 1 SYSTEM 1E Piping Schematic ...
Page 121: ...12 764333 676 Figure 2 SYSTEM 1E Complete 4 3 2 ...
Page 128: ...19 764333 676 Figure 5 Fluid and Air Components 2 of 2 51 53 48 47 50 49 52 ...
Page 133: ...24 764333 676 Figure 7 Float Block Switch Assembly 1 5 8 7 5 4 10 4 2 6 9 9 3 CK2 CK3 LS3 CK1 ...
Page 137: ...28 764333 676 Figure 9 Drain Block Assembly 3 4 5 6 7 3 1 2 8 CK8 TC3 ...
Page 143: ...34 764333 676 Figure 12 Bulkhead Connector Assembly 1 2 2 2 3 1 1 3 5 4 4 6 ...
Page 155: ...46 764333 676 Figure 18 Harness Cable Assemblies 1 of 2 1 2 3 4 5 6 P10 PT1 P5B P11 P2 P5A ...
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......