2-1
764333-675
2.1 GENERAL
2.1.1 Variable Names
General:
During the DIAGNOSTIC Cycle operation of
the SYSTEM 1E™ Liquid Chemical Sterilant
Processing System, the unit measures and stores the
following values into memory. These values can be
accessed through Service Mode to assist in
troubleshooting.
Except where noted, all variables apply to the
DIAGNOSTIC Cycle only. The variables, names, and
purposes are:
2.1.2 How To Access Service Mode
To access Service Mode:
1. Turn Control Power switch OFF.
2. Turn ON Control Power while holding the 3 key on
keypad.
3. Continue to hold 3 key until the unit beeps. Display
shows the following:
ENTER SERVICE
ACCESS CODE 0000
4. The access code is the current year. Enter access
code using keypad, and press
SAVE VALUES
button.
SERVICE OPERATE
BURN IN
From above menu, following choices are available:
• Select
SERVICE
to access test routines and
parameter adjustments.
• Select
OPERATE
to return unit to a normal
operating mode, but with ability to display cycle
parameters while running.
• Select
BURN-IN
for RAM to be cleared and default
parameters to be reloaded.
NOTE: The information on the printout is related to the
software version currently on the control board in the
processor.
NOTE: If MaxPure Filter installed more than 90 days,
a warning print out occurs:
Table 2-1. Variable Names
Variable
Name
Purpose
FZ
Fill Zero
Zero pressure with the
chamber full.
FP Fill
Pressure
Pressure
across
MaxPure™ Filter during
fill.
CZ
Circulation
Zero
Zero pressure with
chamber full, circulation
pump on, and
circulation pinch valve
closed.
CP
Circulation
Pressure
Pressure with the
chamber full and the
circulation pump on.
BT
Begin
Temperature
Starting temperature of
heat rate test.
PS
Pressure Drop
Pressure drop across
membrane of the
MaxPure filter during
filter integrity test.
W A R N I N G
MAXPURE FILTER
IN USE 90 DAYS
SEE OPERATOR MANUAL
Figure 2-1. MaxPure Filter Alert Printout
SERVICE MODE
PROGRAM: 93919091
REV. X XX/XX/XX
CRC = X X X X
BOOT VERSION 6001
XX/XX/XX X:XX:XX PM
RUN TIME = X.X.X hr
CYCLE CNT = XX
serial no. = xxxxxxxxxx
Figure 2-2. Service Mode Information Printout
Section 2: Service Mode Procedures
Summary of Contents for SYSTEM 1E
Page 1: ...MAINTENANCE MANUAL SYSTEM 1E Liquid Chemical Sterilant Processing System 12 09 10 P764333 674 ...
Page 3: ......
Page 71: ...5 19 764333 675 Figure 5 2 Analog Cable Assembly Pin Out Locations ...
Page 74: ...5 22 764333 675 Figure 5 7 DC Power Supply Harness Pin Out Locations PS P1 ...
Page 96: ...7 19 764333 675 Figure 7 17 Aquafine STERIUV5 Control Panel The Control Panel A B 3 2 1 ...
Page 108: ......
Page 110: ...ii 764333 676 Example 1 How To Use IPB Page Typical ...
Page 119: ...10 764333 676 Figure 1 SYSTEM 1E Piping Schematic ...
Page 121: ...12 764333 676 Figure 2 SYSTEM 1E Complete 4 3 2 ...
Page 128: ...19 764333 676 Figure 5 Fluid and Air Components 2 of 2 51 53 48 47 50 49 52 ...
Page 133: ...24 764333 676 Figure 7 Float Block Switch Assembly 1 5 8 7 5 4 10 4 2 6 9 9 3 CK2 CK3 LS3 CK1 ...
Page 137: ...28 764333 676 Figure 9 Drain Block Assembly 3 4 5 6 7 3 1 2 8 CK8 TC3 ...
Page 143: ...34 764333 676 Figure 12 Bulkhead Connector Assembly 1 2 2 2 3 1 1 3 5 4 4 6 ...
Page 155: ...46 764333 676 Figure 18 Harness Cable Assemblies 1 of 2 1 2 3 4 5 6 P10 PT1 P5B P11 P2 P5A ...
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......